Compensation Of Test Leads Resistance - Sonel MIC-2501 Operating Manual

Insulation resistance meter
Table of Contents

Advertisement

Additional information displayed by the meter
NOISE!
R
CONT
in red and two-
tone acoustic sig-
nal is generated

4.2.2 Compensation of test leads resistance

In order to eliminate the impact of the resistance of test leads on measurement result R
compensation (auto-zeroing) of resistance may be performed.
In order to remove the compensation of the leads resistance (return to default cal-
ibration), perform the above-mentioned activities with test leads open, instead of
test results, the display will show message oFF (compensation of test leads is
turned off).
16
Interference voltage occurs on the tested object. The
measurement is possible however it will be burdened
with additional uncertainty that is specified in the tech-
nical data.
, LED
is blinking
Interference voltage exceeds the allowable value, the
measurement is blocked.
In R
ter the screen with Autozeroing the test leads.
Return to R
SET/SEL
OPERATING MANUAL MIC-2501 version 1.04
mode (LED
CONT
Press START.
measurement screen by pressing
CONT
is on) use SET/SEL button to en-
Short the test
leads – message
READY will be
displayed.
AUTO-ZERO message starts to
blink, which confirms completion
of test leads calibration.
The result is a compensated
value and correction is available
for R
. The compensation is
CONT
active even after the meter is
switched off and on again.
, the
CONT