Sonel MIC-10s1 Operating Manual

Sonel MIC-10s1 Operating Manual

Insulation resistance meter
Hide thumbs Also See for MIC-10s1:
Table of Contents

Advertisement

Quick Links

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MIC-10s1 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Sonel MIC-10s1

  • Page 5 OPERATING MANUAL INSULATION RESISTANCE METER MIC-10s1 SONEL TEST & MEASUREMENT Santa Clara, Ca 95051 USA SONEL SA Świdnica, Poland Version 1.03...
  • Page 6 The MIC-10s1 is a high-quality testing instrument, designed for safe and easy operation. Please ac- quaint yourself with this manual in order to avoid measuring errors and prevent possible problems in operation of the meter. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
  • Page 7: Table Of Contents

    Measurement of resistance of protective conductors and equipotential bonding with ±200 mA current ..............33 3.3.2 Calibration of test leads ................35 ................36 EMPERATURE EASUREMENT MEMORY OF MEASUREMENT RESULTS ............ 37 ................37 TRUCTURE OF THE EMORY OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 8 Additional uncertainties according to EN 61557-2 (R ) ......61 10.2.2 Additional uncertainties according to EN 61557-4 (R ) ......61 CONT EQUIPMENT ......................62 11.1 ..................62 TANDARD EQUIPMENT 11.2 ..................62 PTIONAL ACCESSORIES MANUFACTURER ....................64 OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 9: Safety

    This charge must be discharged after measurements to avoid to the risk of electric shock. The MIC-10s1 will discharge the electric charge at the end of the test, automatically.
  • Page 10: Menu

    Wireless transmission  Measurement Settings  Meter Settings  Language selection  Information about manufacturer buttons to select desired po- sition. Enter a selected option by pressing ENTER. 2.1 Wireless transmission See sections 5.3 to 5.5. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 11: Measurement Settings

    The meter is designed to filter out interference generated by 50 Hz and 60 Hz networks. You must choose the correct frequency settings for optimum performance. to select the mains frequency. Confirm your choice by pressing ENTER. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 12: Times T1, T2, T3 For Calculating Absorption Coefficients

    ), t (1 s to 600 s, but >t 2.2.3 Type of absorption coefficients buttons to select the type of co- efficients: Ab or DAR PI. Confirm your choice by pressing ENTER. 2.2.4 I test current OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 13: Setting Limit Values

    Confirm your choice by pressing ENTER. 2.2.6 Temperature and distance units buttons to scroll between dis- tance and temperature and buttons to select the unit. Confirm your choice by pressing ENTER. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 14: Auto-Incrementing Of The Memory Cell Number

    Confirm your choice by pressing ENTER. 2.2.9 Chart type selection To graphically present the measurement results you may select either: current and resistance (I, R) plots, or voltage and resistance (V, R) plots. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 15: Meter Settings

    LCD contrast  Automatic shut-off  Date and time  Factory (default) settings  Firmware update  Button sounds  Wireless communication buttons to select desired po- sition. Enter a selected option by pressing ENTER. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 16: Lcd Contrast

    Select contrast value using buttons. Confirm choice by pressing ENTER. 2.3.2 Automatic shut-off (Auto-OFF) Changing the shut-off time of an idle meter. buttons to set Auto-OFF time. Confirm choice by pressing ENTER. 2.3.3 Date and time OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 17: Factory (Default) Settings

    Before updating the meter’s firmware download the latest version of firmware from the Downloads section at either web-site: www.sonel.pl or www.soneltest.com and save the file on your PC. Connect the meter to PC. Select Firmware upgrade in the MENU and follow the instructions displayed by the program.
  • Page 18: Wireless Communication

    Confirm your choice by pressing ENTER. 2.3.7 Wireless communication buttons to switch on/off wireless communication. Confirm your choice by pressing ENTER. 2.4 Language choice buttons to select desired lan- guage and press ENTER. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 19: Information About Manufacturer

    2.5 Information about manufacturer OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 20: Measurements

    The current limitation occurs in the first phase of the measurement due to charging the capacity of the tested object. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 21: Double-Lead Measurement

    The meter measures the interference voltage V of the tested object. To change the measuring voltage, press F1 buttons to set a voltage value. Confirm choice by pressing ENTER. To set the measurement time, press F3 OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 22 10 nF to 990 nF. When setting to --- (below 10 nF or above 990 nF) the function to calculate the length is switched off. In order to set the limit (minimum resistance), press F4 and ENTER buttons to enter the resistance value. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 23 Value of interference voltage can be read on the display. Connect test leads according to the following diagram: Press and hold START button. The measurement is performed continuously until the START button is released or the pre-set time is reached. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 24 Pressing F4 while in chart view will return to a tabular view of the results. When the measurement is completed the result is displayed. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 25 - For power cables measure the insulation resistance between each conductor and other conductors shorted and grounded (figure below). - The length of the cable is calculated from the value of capacity in [km] entered in the menu before the measurement. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 26: Three-Lead Measurement

    In order to eliminate the influence of surface resistance in transformers, cables, etc. the three-lead measurement is used, but do not connect the current measuring test lead R - to large ground conduc- tors. For example: OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 27 G socket of the meter as follows: (Especially important in humid or wet weather conditions.) OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 28 G terminal to ignore other conductors that do not take part in the measurement.  For the measurement of insulation resistance of a high voltage circuit breaker, the G socket of the meter should be connected to the insulators of the breaker: OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 29: Measurements With Autoiso-5000

    (3-, 4- or 5-wire lead). buttons to select an appropriate item and confirm it by pressing ENTER. buttons, to enter the second group of param- eters. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 30 It is the same for all conductor pairs of a cable. and ENTER buttons to enter the resistance value. and ENTER buttons to select the unit. Confirm by pressing F5 OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 31 If any of the voltages exceeds allowable voltage, the symbol “!” is displayed (e.g. V !), and the measurement ceases. N-PE View of the screen during measurement. After the measurement is completed the results are displayed. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 32: Measurements With Increasing Voltage - Sv

    3.2.2 Press and hold START button. The measurement is performed continuously as long as the button is held or till the end of the measurement. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 33: Dielectric Discharge Indicator - Dd

    Initially this current is the sum of the capacitance discharge current - which fades quickly - and the absorption current. The leakage current is negligible because there is no test voltage. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 34 3.2.2 Start the measurement similarly as in sec- tion 3.2.4. Either during, or after the measure- ment, the user can switch between the text display of the results and the chart by pressing F4 OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 35 - In the Dielectric Discharge mode test it is also possible to perform the measurement with AutoISO- 5000 adapter. The display of results is similar to the measurement of R with AutoISO-5000. The dsipaly will show the following: OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 36: Damage Location (After-Burning)

    F4 When insulation breakdown occurs the meter does not interrupt the measurement (as in other functions). The symbol is shown at the top of the display. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 37: Low-Voltage Measurement Of Resistance

    3.3.1 Measurement of resistance of protective conductors and equipotential bonding with ±200 mA current Set the rotary switch at the R position. CONT The meter is ready for meas- urement. Press F4 set the maximum resistance value. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 38 Voltage on object >10 V + two- Interference voltage exceeds the allowable value, the tone, continuous measurement ceases. beep + and LED lit in red OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 39: Calibration Of Test Leads

    The compensation values are retained even after the meter is switched off and on again. In order to remove the calibration values made above (return to default calibration) perform the above step with test leads not connected together. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 40: Temperature Measurement

    To ensure user safety, it is not allowed to mount ST-1 temperature probe on objects with voltage higher than 50 V to earth. It is advisable to ground the examined object before mount- ing the probe. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 41: Memory Of Measurement Results

    Successive number / Object symbol total number of objects Symbol of the measuring point and the number of such points (for this object) An object containing at least one sub-object Object symbol on a grey background Connector OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 42 Information on the name Entry field entering mode On-screen keyboard To obtain larger fonts set the cursor on Shift and press ENTER. To obtain special (Polish) fonts set the cursor on ALT and press ENTER. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 43: Storing The Measurement Results In The Memory

    4.2 Storing the measurement results in the memory Press ENTER after finishing the measurement. 4.2.1 Entering the results without extending the memory structure Press ENTER again. The cell is free for a given type of measurement. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 44 After selecting YES press ENTER to overwrite the re- sult or ESC, to cancel. Note: - Complete set of results (main result and supplementary results) for a given measuring function and preset measurement settings are stored in the memory. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 45: Extending The Memory Structure

    4.2.2 Extending the memory structure Press ESC to start creating the objects. Press button to set the cursor on Client 1. buttons to select other clients (1 - 10) . Press F1 to edit client data. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 46 In this way, you may enter all client data. Press F5 to confirm data and return to the screen of step button to set the cursor on the object icon. Press F1 to start edit- ing the object name. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 47 (up to 99 items), and pressing F2 deletes the items. Press F5 to approve the name that appears on the screen. Press ENTER, go to the measurement point. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 48 Enter the name of the measuring point in the same way as the object name. Press ENTER, to save the measurement result. After entering the memory, the user may extend its structure by adding new objects and sub- objects as needed. To add a new object, press F5 OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 49 - Deleting objects and sub-objects is possible only in memory browse mode. - The name of an object, sub-object or measurement is possible in memory browse mode or after en- tering entry into the memory after a measurement. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 50: Viewing Memory Data

    Press F1 to enter the option for editing the name of object (sub-ob- ject) and change it. Press F2 to delete a chosen object (sub-object) with all its results. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 51 Press F4 or F5 to display all individual types of results for a given point. Press F2 to delete a chosen measurement or measuring point with all its results: OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 52: Deleting Memory Data

    4.4 Deleting memory data Set the rotary switch of function selection at MEM position. buttons to select "MEMORY ERASING". Press ENTER button. buttons to select YES or NO. Press ENTER button. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 53: Data Transmission

    To communicate from the meter to a PC a USB cable is required, or Bluetooth adapter, and the software supplied with the meter. The software may be used for many devices manufactured by SONEL S.A. that are equipped with the USB interface.
  • Page 54: Connecting To Bluetooth Mini-Keyboard

    From the list of available devices select a keyboard and press "ENTER -Connect" - the meter will dis- play the progress bar, and a counts of 30 seconds. During this time use the keyboard to enter a PIN code and confirm it by pressing "ENTER". OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 55 - Wireless connection error Wrong PIN number entered - connection failure. The PIN code entered does match with the one set in the meter. - Wireless connection error Device not found - the keyboard is not available for connection. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 56: Automatic Connection

    3. Use buttons to select "BLUETOOTH COMMUNICATION", press ENTER button. 3. On the PC enter Bluetooth connectivity mode, select MIC-10s1 device and establish the connection. 4. If the connection is successful the meter will display the following screen: OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 57: Read-Out And Change Of Pin Code For Bluetooth Connections

    5. Start the software to read/save data (e.g., Sonel Reader, Sonel PE) and proceed in accordance with its instructions. 5.5 Read-out and change of PIN code for Bluetooth connections Select Wireless transmission in the main MENU of the meter, press ENTER.
  • Page 58 Note: Standard PIN code for Bluetooth transmission is "123". OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 59: Power Supply Of The Meter

    6.2 Battery power The MIC-10s1 meter is powered with a lithium-ion which may be replaced only at a designated Sonel servicing center. The charger is installed inside the meter and works only with the manufacturer’s rechargeable battery pack.
  • Page 60: Power Supply From Mains

    - If the electrolyte contained in the Lithium-Ion battery pack, contacts eyes or skin, immediately rinse the affected place with plenty of water and consult a doctor. Protect the battery against unauthorised per- sons and children. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 61: Cleaning And Maintenance

    Before waste equipment is sent to a collection point do not dismantle or disassemble. Observe local regulations concerning the disposal of waste electronics, batteries, and related components. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 62: Technical Specifications

    Measurement of insulation resistance []  1000*V Test voltage accuracy (R [V]): -0+10% of the set value Measurement range acc. to IEC 61557-2 for MIC-10s1 V = 10000V: 10.0M to 40.0T Measurement with DC and increasing voltage (SV) for U = 5 kV...
  • Page 63 For measurement voltages below 100 V the measurement error is not specified.  The cable lenght L is calculated as C/Cx, measurement uncertainty depends on measurement range.  Time Constant TC is calculated as Riso x C, measurement uncertainty depends on measure- ment range. OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 64 ............EN 61326-1:2006 and EN 61326-2-2:2006 ATTENTION! MIC-10s1 meter is classified in terms of Electromagnetic Compatibility (EMC) as instruments of Class A (for use in industrial environments - according to EN 50011). Interference may affect the operation of other devices used in other environments (e.g.
  • Page 65: Additional Data

    Temperature 32°F to 95°F / 0C to 35°C 10.2.2 Additional uncertainties according to EN 61557-4 (R CONT Significant parameter Designation Additional uncertainty Position Supply voltage 0.2% (BAT is not lit) Temperature 32°F to 95°F 0°C to 35°C OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 66: Equipment

    5 m shielded cable, 10 kV, black, (cat. IV 1000V) 1000V)   20 m cable, 10 kV, red, (cat. IV 1000V) 5 m cable, red 10 kV, (cat. IV 1000V), with banana plugs with banana plugs OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 67  10 m cable, 10 kV, red, (cat. IV 1000V), with banana plugs  Carrying case L-7  10 m cable, blue, 10 kV, (cat. IV 1000V), terminated with banana plugs AutoISO-5000 adapter OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 68: Manufacturer

    +1 (408) 988 1346 fax +1 (408) 988 4869 E-mail: office@soneltest.com Web: www.soneltest.com SONEL S.A. Świdnica Poland tel. +48 74 858 38 60 fax +48 74 858 38 09 E-mail: export@sonel.pl Web page: www.sonel.pl OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 69 OPERATING MANUAL MIC-10s1 version 1.03...
  • Page 70 OPERATING MANUAL MIC-10s1 version 1.03...