Fluke 6270A Operator's Manual page 40

Pressure controller/calibrator
Hide thumbs Also See for 6270A:
Table of Contents

Advertisement

6270A
Operators Manual
Item
Slew Rate
Control Limit
(Shown only when
Static is selected)
Stability Limit
(Shown only when
Static is selected)
Ready Tolerance
(Shown only when
Dynamic is
selected. Not shown
in figure )
Safety Limits
Pressure Control
Mode
34
1
2
3
Indicator/Name
The slew rate is a user-configurable control parameter to adjust how
fast the Product controls pressure to the setpoint.
Related and used only with the Static Control mode, the control limit
is used to set an upper and lower pressure limit around the Setpoint.
The Product keeps the pressure within defined user limits. See Static
Control Mode and Control Limits (Only for Static Control).
Related and used only for the Static Control mode, the stability limit
is a configurable tolerance to define the ready condition for the
measurement. The measurement indicator shows "Not Ready" until
the pressure is as stable as the value set in the stability limit. See
Stability Limits (Only for Static Control) for more information.
Related and used only with Dynamic Control mode, the Ready
Tolerance is a configurable band to define the ready condition for the
measurement. The band creates a positive and negative boundary
around the Setpoint pressure that is specified in pressure units or
%range of selected module. The measurement indicator shows "Not
Ready" until the pressure is stable within the band. See Ready
Tolerance for more information
Protective user-configurable limits to protect the UUT from over-
pressure. An upper, lower, and vent limit can be manually set. See
Safety Limits.
Shows and switches the active control mode between Dynamic and
Static. See Pressure Control Settings.
Table 10. Control Settings
5
4
Function
huo012.eps