Fluke 6270A Operator's Manual page 11

Pressure controller/calibrator
Hide thumbs Also See for 6270A:
Table of Contents

Advertisement

Dimensions
Height.................................................. 147 mm (5.78 in)
Width ................................................... 452 mm (17.79 in)
Depth .................................................. 488 mm (19.2 in)
Rack Mount Dimensions ..................... 3U-19-inch rack
Chassis Pressure Limits
Supply Port ........................................ 23 MPa (3300 psi) gauge
Test Port ............................................ 20 MPa (3000 psi) gauge
Reference Port ................................... 150 kPa (22 psi) absolute
Vent Port ............................................ 150 kPa (22 psi) absolute
Relief Valves
Chassis Supply port relief valve is set to 24.1 MPa (-0/+700 kPa), 3500 psi (-0/+100 psi)
Exhaust port relief valve is set to ~700 kPa (100 psi).
Each PMM includes a module-specific pressure protection device.
Supply Gas Type
Clean Dry N
or Air – Industrial Grade Nitrogen, 99.5 %+
2
Particulate Contamination ................... ≤ 1.25 micrometer (50 microinches)
Maximum Moisture Content ................ -50 °C dew point
Maximum Hydrocarbon Content ......... 30 ppm
Vacuum Supply
>50 liters per minute capacity with Auto Vent feature
Appropriate protections for High Pressure Gauge work system exhaust gas will pass through the Vacuum supply
system.
Interface / Communications
Primary remote Interfaces .................. IEEE, Ethernet, RS232, USB
System Connection ............................. Supports interconnection of 2 or 3 systems
Switch Test Connection ...................... Standard 4 mm Jack:
Aux Drivers ......................................... 4 external Solenoid Drivers
Nominal 24 V dc isolated drive
Maximum 30 V dc w.r.t. chassis ground
24 V dc Drive (Maximum drive 6 W continuous per channel)
Pressure Controller/Calibrator