Page 2
Chillgard LS Diffusion Refrigerant Monitor Chillgard LS Pumped Refrigerant Monitor...
Page 3
MSA Permanent Instrument Warranty 1. Warranty- Seller warrants that this product will be free from mechanical defect or faulty workmanship for a period of two years from date of shipment, provided it is maintained and used in accordance with Seller's instructions and/or recommendations. This...
Page 4
9. Ensure that the Chillgard LS Refrigerant Monitor is installed in a clean and dry area as possible.Install splash or rain shields on the instrument to keep water and other contaminants away from the instrument.
Page 5
10.Ensure that the Chillgard LS Refrigerant Monitor and the area the instrument is monitoring is at the same temperature before operation of the instrument; otherwise, condensation will form and possibly clog or fill the gas sensor in the instrument rendering the instrument inoperable.
Table 1-5. Typical Chillgard LS / R-123 Cross-Sensitivity Response Data ..1-6 Table 1-6. Typical Chillgard LS / R-134 A Cross-Sensitivity Response Data ..1-7 Chapter 2 Installation and Set-up .
Page 7
Figure 4-2. Applying Calibration Gas to the Chillgard LS Refrigerant Monitor ..4-3 Calibration Guidelines ......4-4 Calibration Equipment .
Also, the sensors are not poisoned or degraded in the presence of most silicon or sulfur compounds. Each Chillgard LS Refrigerant Monitor is placed in an area where a refrigerant is anticipated or must be controlled. The Monitor operates in...
This label contains: • model number • the refrigerant that the unit will detect • information concerning proper voltages and frequency required to power the unit Terminology Used in this Manual FLOW RATE: Volume of gas drawn through the sample line per minute. EXHAUST GAS: Sample gas after it passes through the sensor.
It may be used for applications requiring long-term stability and low maintenance. Applications The Chillgard LS Refrigerant Monitor is factory-calibrated for a refrigerant and can be used for applications such as: • ventilation control •...
Table 1-1. General Operating Specifications for all Models PERFORMANCE SPECIFICATIONS LINEARITY 20 to 100 ppm +5 ppm 100 to 1000 ppm +6% of reading REPEATABILITY +8 ppm WARM-UP TIME 10 minutes RESPONSE TIME 50% of a step-change in < 70 seconds OPERATING TEMP.
• Remove the plastic screws holding the cover to the enclosure. • Remove the front cover from the Chillgard LS Refrigerant Monitor to expose the electrical printed circuit board. Initial Inspection With the front cover open, carefully inspect the components and assemblies inside the enclosure.
1. Most refrigerants are heavier than air and sink to the floor or accumulate in low areas. 2. Place the end of the sampling line in an area that provides the instrument with a representative sample; ventilation Smoke Tubes (P/N 458480) are useful in determining air flow patterns in ambient sampling areas.
Do not locate these units in an area that may contain a flammable mixture of gas and air; otherwise, an explosion may occur. Install the Chillgard LS Refrigerant Infrared Monitor in the area where a refrigerant is anticipated or control of the gas is desired.
1. Open the unit by removing the plastic screws securing the lid to the enclosure. 2. Using #10 mounting screws or bolts (FIGURES 2-1 and 2-2), securely mount the Chillgard LS Refrigerant Monitor to a wall or flat mounting surface via the four holes that run through the enclosure base.
Wiring The Chillgard LS Refrigerant Monitor is factory-wired with either a universal power supply for 110/220 VAC operation or no power supply for 24 V AC/DC operation. Refer to FIGURE 1-1 to determine power requirement for the unit.
Figure 2-5. Sensor Board Connections For 24 V DC or AC Wiring NOTE: The 24-volt power source used with this equipment must be separated from mains by double or reinforced insulation. All power wiring is connected to the terminal block for 24 V DC/AC input shown in FIGURE 2-5.
Analog Signal Output Wiring The Chillgard LS Refrigerant Monitor is factory-configured with an analog output. The analog output for your Chillgard LS Refrigerant Monitor (FIGURE 1-1) is 4-20 mA, current sourcing type (standard) (4 mA = 0 ppm gas, 20 mA = 1000 ppm gas).
Increase the calibration frequency when any calibration differs as much as 10% from the test concentration. More frequent calibrations may be required when the Chillgard LS Refrigerant Monitor is new. Calibration frequency depends on the operating time and chemical exposures of the instrument.
Page 27
If this calibration procedure cannot be performed at any step: • See Section 5, "Troubleshooting" • Localize the problem • Replace the inoperative component. Calibrate newly installed instruments on a frequency until calibration records prove instrument stability. Calibration frequency is then reduced in accordance with a schedule established by the safety officer or facility manager.
Both ZERO gas and SPAN gas must be carefully applied to the unit to avoid pressurizing the internally mounted optical bench. See TABLE 5-2, Calibration Accessories Parts List for the appropriate calibration parts and calibration gases available for the Chillgard LS Refrigerant Monitor. " WARNING Exercise care during the span calibration to ensure that the unit can accurately detect refrigerant gas.
2. Deactivate the equipment connected to the outputs, or disconnect the wiring to the outputs. " CAUTION If any control instruments connected to the Chillgard LS Refrigerant Monitor are wired to external devices (e.g., horns, exhaust fans, and fire suppression systems), these devices may activate while adjustments or repairs are performed during the following procedures.
Figure 4-2. Applying Calibration Gas to the Chillgard LS Refrigerant Monitor 2. Using a multimeter to read the 4-20 mA output, use the Calibration Box to move the zero up or down, until the unit output reads 4 mA +0.1 mA with the zero gas applied.
Keep written records of the calibration readings obtained and any adjustments made. Analysis of these records enables review and control of the time between checks. Check a new Chillgard LS Refrigerant Monitor installation at least once a week by performing the steps outlined in the following section. Calibration Equipment Calibration of the monitor requires a supply of: •...
1. Connect the mA current meter to the instrument output, terminals (+) and (-). " CAUTION During calibration, the Chillgard LS Refrigerant Monitor is not sampling and monitoring the intended area. Exercise caution in the area as appropriate. 2. Connect the ZERO gas cylinder to the sample tubing.
9. Re-connect or enable all equipment and alarm devices connected to any control equipment monitoring the Chillgard LS Refrigerant Monitor. 10.Reinstall the light gray cover. " CAUTION Do not leave any alarm device or equipment disabled or disconnected during normal operation of the instrument; otherwise, the instrument will not function as intended when it detects an alarm situation.
Figure 4-3. Kit Components The Calibration Kit contains a ZERO gas scrubber which can be used in place of a ZERO gas cylinder if the ambient air around the Chillgard LS contains little or no refrigerant. Relative humidity may have a small effect on the output of the unit. If dry gas is used, Nafion Tubing (P/N 813628) can be used to humidify the sample stream going to the monitor.
2. Deactivate the equipment connected to the outputs, or disconnect the wiring to the outputs. "! CAUTION If any control instruments connected to the Chillgard LS Refrigerant Monitor are wired to external devices (e.g., horns, exhaust fans, and fire suppression systems), these devices may activate while adjustments or repairs are performed during the following procedures.
7. Remove all calibration accessories from the unit to return to normal operation. Calibration Guidelines Once the Chillgard LS Refrigerant Monitor is operating, perform periodic calibration checks to ensure proper instrument operation. Perform calibration to monitor long-term changes (drift) in both the ZERO and SPAN readings.
Keep written records of the calibration readings obtained and any adjustments made. Analysis of these records enables review and control of the time between checks. Check a new Chillgard LS Refrigerant Monitor installation at least once a week by performing the steps outlined in the following section. Calibration Equipment Calibration of the monitor requires a supply of: •...
1. Connect the mA current meter to the instrument output, terminals (+) and (-). "! CAUTION During calibration, the Chillgard LS Refrigerant Monitor is not sampling and monitoring the intended area. Exercise caution in the area as appropriate. 2. Connect the ZERO gas cylinder (if used) to the sample tubing.
8. Remove tubing from the sensing cell on the unit. 9. Re-connect or enable all equipment and alarm devices connected to any control equipment monitoring the Chillgard LS Refrigerant Monitor. 10.Reinstall the light gray cover. " CAUTION Do not leave any alarm device or equipment disabled or disconnected during normal operation of the instrument;...
Chapter 5 Service and Replacement Parts Troubleshooting Guidelines The Chillgard LS Refrigerant Monitor is designed to provide long and trouble-free monitoring. If repairs are indicated, it is possible the user can diagnose and correct the problem using the following: •...
Table 5-1.Troubleshooting Guidelines (cont.) SYMPTOM SOLUTION Noisy Output Check input power Check tubing connection between optical bench and sound dampening element Replace optical bench Calibration has Significantly Changed from Previous Reading Check that gas is properly applied Clear inlet of particle matter Replace optical bench Pump Not Running Cycle the power and observe the pump;...
Service and Assistance When ordering replacement parts or to obtain assistance regarding any problem with the Chillgard LS Refrigerant Monitor, please provide the following information (found on a label located on the side of the instrument): • serial number •...
Appendix B RS-485 Output The sensor module uses an RS-485 serial interface with Modbus protocol. To communicate to the RS-485 output via a computer interface, an isolated RS-485 converter (such as P/N 10014359) should be used. The baud rate is 19,200 baud with RTU format sent. Each byte has eight bits with no parity and two stop bits.
Table B-2. Sensor Unit Register Addresses TITLE REGISTER READ/ DEFINITION OF ADDRESS WRITE ENCODED DATA Fault & Module State 40001 Cal Sum 1 msb Under Range 2 msb Flow Loss 4 msb Under Voltage 8 msb Temperature 10 msb Prog Cksum 20 msb Lamp Fail 40 msb...
Need help?
Do you have a question about the Chillgard LS and is the answer not in the manual?
Questions and answers