AC Charger

Symptom
The green and orange-yellow CHARGER ON indicators
illuminate alternately.
Unit displays the LOW BATTERY message when it is
plugged into AC mains.
None of the CHARGER ON indicators are illuminated
when the device is plugged into AC mains.
9650-1210-01 Rev. T
Troubleshooting Guidelines
Recommended Action
Verify battery is installed.
Turn unit ON to identify the fault condition.
Replace battery pack with a fully charged battery pack.
If problem persists, replace battery pack, unplug device
from AC mains and plug device back into AC mains.
Replace battery pack with a fully charged battery pack.
Unplug device from AC mains and plug device back into
AC mains.
Verify AC mains is working properly.
Unplug device from AC mains and plug device back into
AC mains.
Verify AC mains is working properly.