Standard apex/sternum adult and pediatric. 
Paddles
Adult paddles slide off to expose pediatric paddles.
Specifically designed adult anterior/posterior pregelled ZOLL MFE Pads and 
Multi-Function
Electrode (MFE)
Multi-Function stat-padz packaged in pairs.
Pads
Built-in Defibrillator
Provides verification of the defibrillator charging and discharging without removing paddles
Tester
from storage wells or verification of unit configured with Multi-Function Cable.
Multi-Function
Range:
Electrode
Accuracy: ±10% or 5 ohms, whichever is greater.
Impedance
Measurement
Defibrillation
Evaluates electrode connection and patient ECG to determine if defibrillation is required.
Advisory
Shockable Rhythms: Ventricular fibrillation with amplitude > 100 V and wide complex
ventricular tachycardia with rates greater than 150 bpm. Refer to the ECG Analysis Algorithm
Accuracy section for sensitivity and specificity performance.

ECG Monitoring

Patient Connection
3-lead ECG cable, 5-lead wire ECG cable, paddles or MFE Pads. Selectable by front panel
switch.
Input Protection
Fully defibrillator-protected. Special circuit prevents distortion of ECG by pacer pulse. (Pacer
version only).
Implanted
Dedicated circuitry detects most implanted pacemaker spikes and provides standard display
Pacemaker Spike
marker of spike on ECG trace.
Display
Implanted Pulses
±2mV to ±700mV, 0.1ms to 2ms width, with a recharge constant of 0 to 100ms.
Detected
Note: The pacemaker pulse rejection capability for the E Series with pacemaker pulses
Bandwidth
0.67 Hz - 21 Hz (-3 dB) standard/0.05 Hz - 150 Hz Diagnostic
0.67 Hz - 27 Hz and 1 Hz - 21 Hz user-configurable
Lead Selection
Displayed on monitor
ECG Size
0.5, 1, 1.5, 2, 3 cm/mV (Centimeter per millivolt) display on monitor
Heart Rate
Digital display
Range:
Accuracy: ±5%
9650-1210-01 Rev. T
0 ohms - 250 ohms
alone includes pulses between ±2mV and ±700mV amplitude, with widths between
0.1ms and 2ms, without overshoot, and between ±100mV and ±500mV, with widths
between 0.1ms and 2ms, with overshoot from 4 to 100ms.
The pacemaker pulse rejection capability for the E Series with pacemaker pulses and a
normally paced QRS and T wave includes pulses between ±2mV and ±700mV
amplitude, with widths between 0.1ms and 2ms, without overshoot, and between ±2mV
and ±500mV, with widths between 0.1ms and 2ms, with overshoot from 4 to 100ms.
The pacemaker pulse rejection capability for the E Series with pacemaker pulses with an
ineffectively paced QRS pattern includes pulses between ±100mV and ±700mV
amplitude, with widths between 0.1ms and 2ms, without overshoot, and between
±100mV and ±500mV, with widths between 0.1ms and 2ms, with overshoot from 4 to
100ms.
The E Series is not capable of rejecting A-V Sequential pacemaker pulses.
0 bpm - 300 bpm 
A-3