Screen icon indicates activated/deactivated status. User-selectable. 
Heart Rate Alarm
Tachycardia 60 bpm - 280 bpm, bradycardia 20 bpm - 100 bpm
Tall T-wave
1.0 mV
Rejection
Heart Rate
The E Series averages the interval between the last 5 detected beats. On startup, the
Averaging
E Series averages the rate between detected beats once two beats are detected, until a full 5
beats have been received. The rate is updated every beat. After this condition is met, the
meter is updated every beat with an average of the last 5 beats.
If a period of time greater than 5 seconds elapses without a beat detected, the heart rate
meter reports a rate of 0 bpm, which is repeated every 5 seconds.
Accuracy and
Averaging over 5 R-R intervals, per AAMI EC 13:2002:
Response Time to
Ventricular Bigeminy (Figure 3a) - 40 bpm
Irregular Rhythm
Slow alternating ventricular bigeminy (Figure 3b) - 60 bpm
Rapid alternating ventricular bigeminy (Figure 3c) - 120 bpm
Bidirectional systoles (Figure 3d) - 90 bpm
Response Time to
80 to 120 bpm: 4 seconds
Change in Heart
80 to 40 bpm: 4 seconds
Rate
Time to Alarm for
206 bpm (1 mV): 4.80 seconds
Tachycardia
206 bpm (halved amplitude): 9.90 seconds
206 bpm (doubled amplitude): 4.00 seconds
195 bpm (2 mV): 4.20 seconds
195 bpm (halved amplitude): 8.00 seconds
195 bpm (doubled amplitude): 4.00 seconds
Leads Off Sensing
A DC current of 0.04
Active Noise
The sum of all leadwire currents is returned via the active noise suppression leadwire:
Suppression
0.08
0.16
0.36
1 Volt ECG Out
1.0 Volt/cm (volt per centimeter) of deflection on stripchart recorder
< 25 ms delay from patient ECG input
Bandwidth 
Output impedance = 250 ohms
Internal pacemaker pulses represented as on display
ECG x 1000 provided at rear connector panel via a standard 3.5 mm mono jack. ECG signal
is on tip; ground is on ring
Display Format
Non-fade moving bar display

CPR Monitoring

Compression Depth 0.75 to 3 inches ±0.25 inches
1.9 to 7.6 cm ±0.6 cm
Compression Rate
50 to 150 compressions per minute
A-4
A per lead wire is supplied to the patient.
A DC in 3 lead mode
A DC in 5 lead mode
A DC in 12 lead mode
150 Hz
9650-1210-01 Rev. T