ZOLL E Series Operator's Manual page 113

Defibrillator
Hide thumbs Also See for E Series:
Table of Contents

Advertisement

E Series Operator's Guide
Table A-1.
Delivered Energy at Every Defibrillator Settings into a Range of Loads
Selected
Energy
25
1 J
1 J
2 J
1 J
3 J
2 J
4 J
3 J
5 J
4 J
6 J
4 J
7 J
5 J
8 J
6 J
9 J
7J
10 J
7 J
10 J
15 J
11 J
15 J
20 J
15 J
19 J
30 J
22 J
29 J
50 J
37 J
48 J
70 J
53 J
69 J
85 J
65 J
86 J
100 J
74 J
97 J
120 J
88 J
116 J
150 J
110 J
145 J
200 J
146 J
214 J
Figures A-1 through A-20 show the rectilinear biphasic waveforms that are produced when the E Series defibrillator is
discharged into loads of 25, 50, 75, 100, 125, 150, and 175 ohms at each energy setting (200, 150, 120, 100, 85, 70,
50, 30, 20, 15, 10, 9, 8, 7, 6, 5, 4, 3, 2 and 1 joule[s]).
The vertical axis shows the current in amperes (A); the horizontal axis shows the duration in milliseconds (ms).
A-14
Load
50
75
100
1 J
1 J
1 J
2 J
3 J
2 J
3 J
4 J
4 J
4 J
5 J
5 J
5 J
6 J
6 J
6 J
8 J
7 J
7 J
9 J
9 J
8 J
10 J
10 J
9 J
11 J
11 J
13 J
12 J
19 J
19 J
25 J
25 J
38 J
37 J
63 J
62 J
90 J
89 J
112 J
110 J
126 J
125 J
151 J
149 J
188 J
186 J
238 J
234 J
125
150
175
1 J
1 J
1 J
2 J
2 J
2 J
3 J
3 J
3 J
5 J
4 J
4 J
6 J
5 J
5 J
7 J
6 J
6 J
8 J
8 J
7 J
9 J
9 J
8 J
10 J
10 J
9 J
12 J
11 J
10 J
17 J
16 J
15 J
23 J
22 J
20 J
35 J
32 J
30 J
58 J
54 J
50 J
83 J
77 J
71 J
103 J
96 J
89 J
116 J
108 J
100 J
139 J
129 J
120 J
174 J
161 J
150 J
213 J
195 J
179 J
Accuracy
±3J
±3J
±3J
±3J
±3J
±3J
±3J
±3J
±3J
±3J
±3J
±3J
±15%
±15%
±15%
±15%
±15%
±15%
±15%
±15%
9650-1210-01 Rev. T