Rev. b4 high accuracy automated pressure controller
Table of Contents

Advertisement

Command
Type
GO
W
CN
W
IE
R
IE=n
W
HC
R
HC=n
W
C1
R
C1=n
W
C2
R
C2=n
W
C3
R
C3=n
W
C4
R
C4=n
W
CC
W
SF=0
R
SF=1
W
SG
W
GG
W
F3
W
F2
W
Description
Command to execute current settings.
Command to cancel current settings
Read current IEEE address
Set IEEE address to n
Read current Height Correction
Set Height Correction to n
Read Pt vacuum calibration point
Set Pt vacuum calibration point
Read Pt full-scale calibration point
Set Pt full-scale calibration point
Read Ps1 and Ps2 vacuum calibration point
Set Ps1 and Ps2 vacuum calibration point
Read Ps1 and Ps2 full-scale calibration point
Set Ps1 and Ps2 full-scale calibration point
Activate new calibration points after new values have
been entered.
Read the status of the Self Test last performed.
Start the Self Test.
Set 'Ground'
Go To 'Ground'
Change the main screen to 3-output format
Change the main screen to 2-output format