When the last point in the profile is executed, pressing ENTER will cause line 4 of the
display to be cleared. This will be the only indication that the profile has been
completed.
8.
Each profile can have up to 50 points. There can be 12 profiles resident in the 6600 at
any one time.
9.
New profiles can be loaded any number of times. They have to be numbered 1 to 12
and will overwrite existing resident profiles with the same profile number.
10.
Errors during profile execution will cause the profile to abort. It can be restarted by the
user at the point where the error occurred, after the error is rectified.
11.
If a particular test procedure requires more than 50 points, then these points can be
split between 2 or more profiles. For instance, if a transponder test requires 80 points,
then the first 50 points can be loaded as profile 1 and the remaining 30 points can be
loaded as profile 2. When profile 1 has been completed at point 1.50, the 6600 will stay
at that point until the next command. The user can start profile 2.1 and execute the 51st
point and continue through the remaining 30 points using profile 2.