Laversab 6600 User Manual page 70

Rev. b4 high accuracy automated pressure controller
Table of Contents

Advertisement

Model: 6600
Full scale:
Ps: 38 inHg
Pt: 60 inHg
Last Calibrated date:
_______________________________________________________________________
TEST POINT
1. Vacuum
2. Fullscale
TEST POINT
1. Vacuum
2. Fullscale
TEST POINT
1. Vacuum
2. Fullscale
_______________________________________________________________________
Ps1 Output
APPLIED
DISPLAYED
inHg
inHg
1.000
1.000
4.000
4.000
8.000
8.000
12.000
12.000
15.000
15.000
19.000
19.001
23.000
23.001
27.000
27.000
30.000
30.000
34.000
34.000
38.000
38.000
_______________________________________________________________________
Calibrated by:
June 25, 2007
Next Calibration due: June 25, 2009
CALIBRATION RESULTS
Ps1 Output
AS FOUND
inHg
1.001
38.002
Ps2 Output
AS FOUND
inHg
1.000
38.001
Pt Output
AS FOUND
inHg
1.001
60.002
VERIFICATION RESULTS
Ps2 Output
APPLIED
DISPLAYED
inHg
inHg
1.000
4.000
8.000
12.000
15.000
19.000
23.000
27.000
30.000
34.000
38.000
Figure 6.1 Sample calibration sheet
63
Serial#: 77123
AS LEFT
inHg
1.000
38.000
AS LEFT
inHg
1.000
38.000
AS LEFT
inHg
1.000
60.000
0.002
APPLIED
inHg
1.000
1.000
4.000
6.000
7.999
12.000
11.999
18.000
15.000
24.000
19.000
30.000
23.000
36.000
27.001
42.000
30.000
48.000
34.000
54.000
38.000
60.000
Date: June 25, 2008
DEVIATION
inHg
0.001
0.002
DEVIATION
inHg
0.000
0.001
DEVIATION
inHg
0.001
Pt Output
DISPLAYED
inHg
1.000
5.999
11.999
18.000
24.000
30.000
36.000
42.001
48.002
54.001
60.000