Calibration; Equipment; General Notes; Section 6: Calibration - Laversab 6600 User Manual

Rev. b4 high accuracy automated pressure controller
Table of Contents

Advertisement

6.1

EQUIPMENT:

This calibration procedure requires the following equipment:
1. Vacuum pump.
2. Pressure supply of clean, dry air ("Zero" Air preferred) at 35 psia
3. Pressure standard having accuracy of at least 0.002 inHg up to 38 inHg and an
accuracy of at least 0.003 inHg from 38 inHg to 60 inHg.
6.2

GENERAL NOTES:

The transducers inside the 6600 are sensitive to the change in density of the medium being
used. The unit must be calibrated with Dry Air only. Calibrating with Nitrogen or any
medium other than Dry Air will give erroneous results.
Because the transducers are medium-sensitive, care must be taken to ensure that the hoses
used in the calibration system do not out-gas. Normally, PVC hoses, or refrigeration type
hoses have a tendency to out-gas. The ideal hoses to use are Teflon or Teflon-coated hoses
because they have a very low absorption and out-gas coefficient. Stainless steel hoses may
also be used.
If the wrong type of hoses are used, then the typical results obtained will show that the error
increases as the pressure increases. Also, it is very unlikely that the repeatability of the
indicated pressure will be less than 0.002 inHg at 60 inHg. If the system cannot achieve
repeatability of 0.002 inHg or less, then the most likely cause is the out-gassing of hoses.
In the following sections, the instructions call for generating a pressure of 1.000 inHg abs.
using the Pressure Standard. If your Standard cannot generate such a low pressure, then
please generate as low a pressure as possible provided it is less than 4.0 inHg abs.
The 6600 has a drift of typically 0.005 %FS per year. Therefore it needs to be calibrated once a
year to perform adequately as a calibration standard for air data equipment.
SECTION 6

CALIBRATION

58