HIOS HM-10 Operation Manual page 7

The torque value checking meters for automatic controlled system
Table of Contents

Advertisement

5-2. Set the bit end of the target screwdriver to the center of the head screw of Fidaptor.
6. Set the switch for measurement unit change to your desired unit.
POWER
O N
OFF
RESET
7. Set the mode switch to TRACK and turn the zero adjustment knob for zero adjustment.
POWER
O N
OFF
RESET
8. Set the mode switch to PEAK.
POWER
O N
OFF
RESET
◎ Important
Notes when fixing Fidaptor:
(1) In order to have the bit set right in the center of the head screw of
Fidaptor, you need to fix Fidaptor by fastening the four screws on the
side evenly.
(2) The bit tip should be engaged precisely in the head screw of Fidaptor
for measurement (it should be correctly aligned with the thrust
direction and the center of the screw.)
(3) Especially, when you use it in the automatically controlled system, it
should be aligned with the center of the Z-axis.
HM
N m
lbf in
kgf cm
MODE
PEAK
TRAK
MADE IN JAPAN
HM
N m
lbf in
kgf cm
MODE
PEAK
TRAK
MADE IN JAPAN
Turn the knob for zero adjustment.
(Note) Zero adjustment is not available with the PEAK position.
HM
N m
lbf in
kgf cm
MODE
PEAK
TRAK
MADE IN JAPAN
After measurement, clear the value with the RESET button.
Unit converter
Switch
Check if "000" is displayed.
Set the mode switch to TRACK.
Set the mode switch to PEAK.
- 6 -