    However high the quality of your screwdriver, you cannot make full use of its potential unless you know the right amount of torque for each job and make sure that is the torque being delivered. With a HIOS digi- tal torque meter you can make torque settings on screwdrivers, check torque wrenches or make torque measurements of various kinds of rotation tools.
  • Page 3: Safety Precautions

    14. Exceed- ing capacity load levels by 120% will result in internal damage to the meter.  The battery charger is for NiMH batteries only. Nev- er use any other kind of battery, even from HIOS. - 3 -...
  • Page 4: Hios Digital Torque Meter Controls

    Adapter socket For connecting of torque tool, using a HIOS Fidaptor or other type of adapter. Please push the power switch and detach it immediately until a buzzersounds.   * Statistical Process Control (SPC) uses statistical techniques to monitor variations in manufacturing processes for use in quality control.
  • Page 5: List Of Items Included With Torque Meter And Explanations Of Fidaptor And Threaded Shaft

    Supplied Accessories Market Model Fidaptor with Areas Driver Shaft Accessories for exclusive use of HP-100. Safety Guard Fidaptor HIOS shank 4 Hex. shank Grease(*) HIOS shank 5 Buffer Spring (convertible shaft) To use for measuring Korea Lower range: 5 mm subtense Measuring 0.15 −...
  • Page 6: Fidaptor

    (comes with Model HP-100) • Loosen the spring up to the desired area for Use as adapter for bits that are not the 5 mm HIOS storing the Fidaptor properly. standard diameter. (Refer to diagram on page 10 • Scratch or wipe with chemicals such as benzin when changing shock absorber spring.)
  • Page 7: Operating Instructions

    5. Press the 【 RESET 】 switch to cancel the dis- guide table in the operating instructions for play. your HIOS power screwdriver to set the torque 6. Minus signs appear in front of measurements of adjustment nut to an appropriate torque value.
  • Page 8: Explanations Of Functions

    ■ Explanations of functions ting has been changed. The torque setting can be changed as desired. The warning tone signals that DIP switch the set torque value has been measured. Always turn off the power before Example (torque set at 100) changing DIP switch settings.
  • Page 9: Memory

    Memory Fig. 1 Recording readout Objective Example of memory readout Memory mode on (LED on) Measurement data can be stored in the memory for Ready for management of everyday operations. measurement Fit the power screwdriver into the socket, Measurements of driver Storing and reading out measurement data using the Fidaptor or other adapter torque...
  • Page 10: Average Torque Measurement

    Notes Measurement of average torque 1. Be sure to set 【 MODE 】 switch to "PEAK" or Objective "F.PEAK" before entering average measurement mode. Simple monitoring of torque for normal operations 2. The 【 UNIT 】 switch does not function in average Measurement procedure measurement mode.
  • Page 11: Fidaptor Parts

    Item Part No. Note Shaft Range *2 TF6U-Z Fidaptor Hex. shank 3 − 9N•m TF6U-Z For HIOS torque drivers with lever- 35mm 6.35 mm start. 15mm subtense Fidaptor Hex. shank 3 − 9N•m TF6UP-Z For HIOS CL-9000 or high torque 6.35 mm...
  • Page 12: Using The Adapter Socket For Torque Measurement Of Instruments Other Than Screwdrivers

     Do not use the battery charger for any purpose ■ Battery charger other than charging batteries for this instrument. To charge batteries for the HIOS torque meter use  Do not place objects on top of the battery char- only the battery charger provided with the meter.
  • Page 13: Types Of Data Output

    (the HIOS calibration check bar) is provided, but in addition to this, we An analog cord is available for separate purchase recommend that the torque meter be sent in to HIOS ). This cable can be used to 1.5 m, item HP-8060 Corporation for inspection at least once a year.
  • Page 14: Troubleshooting

    ■ Troubleshooting Before you conclude the torque meter is broken, go through the following troubleshooting guide. If, by fol- lowing the instructions, you are unable to resolve the diffi culty, please contact HIOS Corporation or your HIOS dealer. Problem Causes Resolving the problem ...
  • Page 15: External Features Diagram

     As for the dimensions, HP-1, 10 and 100 are the same. Note that specifi cations or the external ap- pearance of HIOS torque meters are sub- ject to change without notice. ■ Analog and data output specifi cations A display of 1000 means approximately 1 volt.