  • Page 1 Hios Digital Torque Meters HP-100 HP-10 Operation manual (November 2005) Operation manual No. ET— C006...
  • Page 2: Table Of Contents

    However high the quality of your screwdriver, you cannot make full use of its potential unless you know the right amount of torque for each job and make sure that is the torque being delivered. With a Hios digital torque meter you can make torque settings on screwdrivers, check torque wrenches or make torque mea- surements of various kinds of rotation tools.
  • Page 3: Safety Precautions

    12. Exceed- ing capacity load levels by 120% will result in internal damage to the meter. The battery charger is for NiMH batteries only. Never use any other kind of battery, even from Hios. - 2 -...
  • Page 4: Hios Digital Torque Meter Controls

    Adapter socket For connecting of torque tool, using a Hios Fidaptor or other type of adapter. * Statistical Process Control (SPC) uses statistical techniques to monitor variations in manufacturing processes for use in quality control.
  • Page 5: List Of Items Included With Torque Meter And Explanations Of Fidaptor And Threaded Shaft

    * Grease is to be applied on the Fidaptor with safety guard and Driver shaft appropriately. Old models of the Fidaptor and the intershank are not compatible with model HP-100. The main unit and accessories come all together in a dedicated aluminum attache case.
  • Page 6: Operating Procedure

    (comes with Model HP-100) 9. It is possible to fashion adapters for torque mea- Use as adapter for bits that are not the 5 mm Hios surement of other types of tools. (See "Using the standard diameter. (Refer to diagram on page 8 when adapter socket for torque measurement of instru- changing shock absorber spring.)
  • Page 7: Explanations Of Functions

    Explanations of functions ting has been changed. The torque setting can be changed as desired. The warning tone signals that DIP switch the set torque value has been measured. Always turn off the power before Example (torque set at 100) changing DIP switch settings.
  • Page 8: Memory

    Memory Fig. 1 Recording readout Objective Example of memory readout Memory mode on (LED on) Measurement data can be stored in the memory for Ready for measurement management of everyday operations. Fit the power screwdriver into the socket, Measurements of driver Storing and reading out measurement data using the Fidaptor or other adapter torque...
  • Page 9: Average Torque Measurement

    (Model HP-100) ments. Two springs are supplied with Model HP-100. The 1. The measurement results are automatically num- black spring is high-tension and the yellow spring bered, from 0001 to 0010. Negative displays are is low-tension.
  • Page 10: Fidaptor Parts

    Hex. shank To use for measuring screwdriver of Hex. shank 3 – 9N•m TF6U-Z guard 6.35 mm subtense 6.35 mm subtense. Standard accessory See the Manual p.4 of Standard accessories TF5G-SET set for HP-100 (without battery charger). - 9 -...
  • Page 11: Using The Adapter Socket For Torque Measurement Of Instruments Other Than Screwdrivers

    +0.060 +0.035 Battery charger Danger! To charge batteries for the Hios torque meter use Leaving the battery charger on for the maximum recharge time of 6 hours could lead to overheat- only the battery charger provided with the meter. ing, damage to the instrument or fires.
  • Page 12: Types Of Data Output

    (the Hios calibration check bar) is provided, but in addition to this, we Repair recommend that the torque meter be sent in to Hios 1. Service charges will be made for repairs under Corporation for inspection at least once a year. (Note the following circumstances: that a fee is charged for this service.)
  • Page 13: Troubleshooting

    Troubleshooting Before you conclude the torque meter is broken, go through the following troubleshooting guide. If, by following the instructions, you are unable to resolve the difficulty, please contact Hios Corporation or your Hios dealer. Problem Causes Resolving the problem The display says "LOBAT"...
  • Page 14: External Features Diagram

    RESET UNIT MODE M_SET M_CR Unit: mm Both Model HP-100 and HP-10 have the same dimensions. Note that specifications or the external ap- pearance of Hios torque meters are subject to change without notice. Analog and data output specifications A display of 1000 means approximately 1 volt.