Bit 5 of Binary Output - Configures the output of channel E to function as Bit 5 in
the Binary output configuration indicating the number (max. 31) of the closest
matching standard within the downloaded project.
Searching – Determines the standard search method. When set to Searching disabled
(default setting), no standard search is used. The active standard is used for sample
comparison. When set to Searching enabled, the instrument is searched for the closest
standard match to the measured sample.
Averaging - Determines the method used to derive an average, along with the number of
measurement taken to obtain a normal or weighted value digitally filtered average, and
how much time elapses between measurements. The Digital Filtering option is an
alternative to simply averaging a set of readings. This method applies a weighting factor
across multiple readings selected. The averaging control can be turned off by setting it to
(0) zero (default setting), or adjusted from 2 to 10 measurements to average. Time
delayed between measurements can be set from 0 to 10 seconds. When the control is set
to 0, the Measure button must be clicked to take the measurement. When set to a time
increment (10 = 1 sec, 20 = 2 sec., etc.), the instrument will pause the defined amount
between measurements.
PLC - Determines the amount of time the instrument will "pause" after seeing the Read
input go activate (discrete input only), and how long the instrument holds its outputs
active after a measurement occurs. The Time between external trigger and reading
control occurs before an actual measurement is taken and can be adjusted from 0 to 10
seconds, in tenths of a second increments (11 = 1.1 sec, 22 = 2.2 sec., etc.). The Time
output control remains active control can be set from 0 to 100 seconds, in tenths of a
second increments (11 = 1.1 sec, 22 = 2.2 sec., etc.).
Use Defaults button
Clicking the Use Defaults button will set all options in the Configuration PLC dialog to
the factory default settings.
Load Settings button
Clicking the Load Setting button will download the current settings in the Configuration
PLC dialog to the Vericolor Spectro instrument.
V E R I C O L O R
®
S P E C T R O
43