Press "SET" to accept the new temperature limit.

8.10.2 Soft Cutout

The Soft Cutout parameter is used by the controller to shut the unit down during
over-temperature conditions.

Press "SET" to enable adjustments of the Soft Cutout.
1225
1225.
Adjust this parameter by using "UP", "DOWN", and "SET" as each digit flashes.
 
Press "SET" to accept the new temperature limit.
If the temperature of the unit is ever greater than the Soft Cutout temperature the
controller shuts itself down and displays, alternately, "SCtOut" and "Err 8".

8.10.3 Cutout Reset Mode

The cutout reset mode determines whether the cutout resets automatically when
the well temperature drops to a safe value or must be manually reset by the
operator.
The parameter is indicated by,

Press "SET" to access the parameter setting. Normally the cutout is set for
automatic mode.

8-12
Find Quality Products Online at:
New HL setting
Soft Cutout parameter
Access Soft Cutout
Flashes the current value and then displays the value for
adjustment
Current Soft Cutout setting
New Soft Cutout setting
Cutout reset mode parameter
Cutout set for automatic reset