4. Retrieve the original calibration errors from the unit or the Report of
5. Record the previous values for CE1, CE2, and CE3 and calculate new values
6. Enter the new CEm value in the calibration parameter menu using either the
11-2
Find Quality Products Online at:
Calibration.
for CE1, CE2, and CE3 using the following formula.
Tmn –Tsn +CEn = CEm
Where:
Tmn is the temperature measured, Tsn (CTn) is the set-point temperature,
CEn is the old value for calibration error, and CEm is the new value for
calibration error.
keypad or through the serial port.