Thermo Fisher Scientific
Response Message
address
xxx
function
06H
add hi
xxx
add lo
xxx
value hi
xxx
value lo
xxx
crc16 lo
xxx
crc16 hi
xxx
Example 6: Host Wants to Write the Batch Number
Batch number = 0
Address of Micro-Tech (Slave) = 1
Host (PLC)
Slave (Micro-Tech)
→
01H
06H
00H
5CH
00H
00H
49H
→
D8H
←
←
Micro-Tech 9101/9201 Reference Manual, Rev J
01H
06H
00H
5CH
00H
00H
49H
D8H
C-35