MAT'L CALIBRATION
3 points linear.
MAT'L CALIBRATION
Thermo Fisher Scientific
12. Press the Yes button to calculate the new span number, and the
following screen appears. (If you press the Advance button, the
Micro-Tech displays the difference from the previous setting.)
READY
BATCH
ALARM
SPAN # CHANGED
New span #
xxxxxx
Old span #
xxxxxx
RUN
MENU
FACTOR
13. Press the Run button to return to the Material Calibration screen
shown below. (If you press the Factor button, the Micro-Tech uses
a simulated calibration to populate the parameters in the Material
Factor function.)
READY
BATCH
ALARM
MAT'L CALIBRATION
Add reference
weight to totals?
YES
NO
14. Press the No button and the Run screen reappears.
Here is the Material Calibration menu table.
Selection
no, yes
START, MENU
CALIB
CALIB
Description
This scroll is only available
when Linearization scroll 1
is set to Yes.
Micro-Tech 9101/9201 Reference Manual, Rev J