Thermo Scientific Micro-Tech 9101 Reference Manual page 179

Belt conveyor scale integrator

Advertisement

Thermo Fisher Scientific
READY
BATCH
ALARM
- MAIN MENU 2 -
Press MENU for more
DISPLAY
SCALE
CALIB
3. Press the Calibration Data button and the following screen appears.
READY
BATCH
ALARM
START OF SCROLL
Use SCROLL keys to
view selections
4. Press the down-arrow button to bring up the Calibration Data screen.
READY
BATCH
ALARM
- CALIB DATA SCROLL 1 -
Calibration mode
> R-Cal <
CHOICE
5. Press the Choice button repeatedly until the method you want to use
appears in the screen, as shown below.
CALIB
CALIB
CALIB
Micro-Tech 9101/9201 Reference Manual, Rev J
B-7