UART Configuration Jumpers
Jumper
J34
J35
J36
J39
J38
J40
J41
J42
J43
Thermo Fisher Scientific
Jumper Settings
COMM A, (UART 0)
Jumper Setting
Description
Pins 1-2
RS-485 Normal Operation
Pins 2-3
RS-485 Multi-Drop Operation
COMM A, (UART 0)
Jumper Setting
Description
Pins 1-2
RS-232 Mode
Pins 2-3
RS-485 Mode
Not Installed
RS-232 Mode
COMM A, (UART 0)
Jumper Setting
Description
RS-485 Receive U0RXD
Pins 1-2
RS-485 Receive U0CTS
Pins 2-3
COMM A Termination, (UART 0)
Jumper Setting
Description
Pins 1-2
Enable RS-485 Termination
Pins 3-4
Disable RS-485 Termination
Not Installed
Disable RS-485 Termination
COMM A Termination, (UART 0)
Jumper Setting
Description
Pins1-2
Enable RS-485 Termination
COMM A Termination, (UART 0)
Jumper Setting
Description
Pins1-2
1.2kΩ termination enable
Pins 2-3
600Ω Bias Enabled
Not Installed
No additional termination
Default
J34 Installed 1-2
J35 Installed Pins 1-2
J36 Installed Pins 1-2
J39 Installed Pins 1-2
J38 Installed 1-2
J40 Installed 3-4
J41 Installed
J42 Installed 1-2
J43 Installed 1-2
Micro-Tech 9101/9201 Reference Manual, Rev J
A-31