High-Level Inputs (Two Channels)
Differential voltage.
Input Range (Volts)
0 to +5 V
+1 to +5 V
-5 to +5 V
Current (Requires Jumper Selection)
Input Range (mA)
0 to +20 mA
+4 to 20 mA
Converted Display (Volts)
0 to +5 V
+1 to +5 V
Jumpers J1 and/or J2 are used to select 250 ohm resistance for the
Current inputs.
Input impedance: 100 k nominal (differential)
Maximum usable input voltage: 106% of full scale
Non-isolated voltage
Max. non-destructive input voltage: 12V peak
Current Outputs (Two Channels)
Here are the specifications.
Optically isolated
Isolated power source
Output Range (mA)
0 to 20 mA
+4 to 20 mA
Voltage output by adding an internal dropping resistor.
Micro-Tech 9101/9201 Reference Manual, Rev J
A-43