Instruction Manual CERATHERM 600-3
The unit meets the requirements of EU Directive 93/42/EEC/Annex II.
Monitoring is provided by TÜV SÜD Product Service GmbH, 80339
0123
München.
The unit must be disposed of in accordance with EU Directive
2002/96/EC. Get in touch with your allocated NUFER MEDICAL sales
partner.
3.

Safety guidelines

Please ensure that the following safety guidelines are observed. The guidelines must be
observed during the installation and use of the radiant heater:
The radiant heater must only be connected to a power socket with a
protective earth connection.
Connection values of the power socket:
230 V AC +/−10%, 50–60 Hz, 6 A.
The unit is fitted with a ceramic heater that is sensitive to knocks and blows.
Ensure that the unit is not exposed to any strong knocks or blows.
Ensure that the wall mount is fastened into solid masonry and that suitable
wall-plug screws are used. Lightweight walls must be appropriately reinforced
in the wall-mount area. The client's specialist personnel are responsible
for the secure fastening of the radiant heater.
-> Risk to user and patient safety.
Parts must neither be removed from or attached to the radiant heater and no
other modifications must be made -> Risk to user and patient safety.
Ensure that there is a clearance of at least 80 cm between the reclining
surface and the bottom edge of the heater head; there must be a clearance of
at least 50 cm between the ceiling/false ceiling and the top edge of the heater
head -> Risk of fire, risk to user and patient safety.
14.171.C_GA_Ceratherm_600_3_EN.docx
Nufer Medical AG / 04.07.2014 / DU
Seite 9 / 32