Instruction Manual CERATHERM 600-3
6.

Product description

The CERATHERM 600-3 radiant heater has four programmable heat output levels, which
can be selected directly by pressing the buttons [heat level 1-4].
The default settings for the heat levels are 20, 50, 75 and 98%. All levels can be adjusted by
a specialist upon request (within certain limits, e.g. safety level no larger than 20%).
All heat outputs above the safety level are monitored with a 15-minute reminder alarm. This
alarm is intended as an aid for the user only; it does not relieve the user of a duty to
supervise the patient.
An indicator lamp is built in to monitor actual heat output.
The patient is lit by a modern high-power LED lamp, which has the following advantages over
a halogen lamp:
• More natural colour temperature (4000°K)
• No UV- or IR-radiation component
• Longer service life
• Lower energy consumption
Observe the relevant information in chapter 15 (Lighting)!
The light can be set to one of two brightness levels (default setting: 20% and 100%). These
values can be set by a specialist in 10% increments.
The lighting switches on and off gradually, i.e. with a smooth transition, so it has a less
disruptive impact on the patient.
In standby mode, the lighting has an automatic shut-off function. If the lighting is switched on
and no lighting button has been pressed for 20 minutes, the lighting switches off
automatically. This function can be disabled as required.
The installed software version can be displayed as necessary.
The setting of the relevant parameters is described in chapters 18 and 19 of the
CERATHERM 600-3 service manual.
14.171.C_GA_Ceratherm_600_3_EN.docx
Nufer Medical AG / 04.07.2014 / DU
Seite 14 / 32