Instruction Manual CERATHERM 600-3
8.

Commissioning

When the radiant heater is switched on via the mains switch on the rear side, the unit is
initially in standby mode.
1. If you require additional light, switch on the LED lamp with a low [8] or high [9]
brightness level.
2. Switch on the heater head by pressing on/off button [7]. The heat output is
automatically set to heat level 1 (20%) to keep the reclining surface warm. (The
reclining surface is pre-warmed after roughly 5–10 minutes.)
3. If greater output is desired as per table 4, select one of the heat levels 2–4 directly
[3,4,5]. The yellow heat output indicator LED [10] displays whether the radiant heater
is providing the desired output.
4. If heat level 2, 3 or 4 is selected, a reminder alarm will sound after 15 minutes. If this
alarm is not confirmed with alarm button [6] within 8 seconds, output is reduced to the
safety level (20%) for safety reasons. If the alarm is confirmed within 8 seconds, the
heater will operate for a further 15 minutes, at the previous heat output.
5. The heater head can be switched to standby mode by pressing the green on/off button
[7].
6. Once the controls have been set to standby mode via on/off button [7] on the front, the
unit can be fully shut down by pressing mains switch [1] on the rear side.
8.1