12
Instruction Manual CERATHERM 600-3
When the housing of the heater head is removed, there is a risk of electric
shock. The unit's mains cable must always be disconnected before the
housing of the heater head is opened.
When the heater head has been on for 15 minutes, an acoustic and visual
alarm is activated. The alarm can be delayed for a further 15 minutes by
pressing the
When the radiant heater is being used above infant incubators or heating beds,
it must be ensured that there is sufficient clearance between the bottom edge
of the heater head and any heat-sensitive materials such as Plexiglas or
acrylic glass. Ensure that there is at least 80 cm clearance at all times. Set the
heat output no higher than level 3 (75%).
The radiant heater must only be operated with original accessories that
NUFER MEDICAL has expressly approved for use with the CERATHERM 600-
3. Observe the relevant information in chapter 16.
In the event of a technical alarm
have it checked by an authorised person and repaired as required. A safety
check must be performed before the unit can be used again.
Before any cleaning or disinfection is performed, ensure that the unit is shut
down and disconnected from the mains supply. Observe the relevant
information in chapter 12.
Repair and maintenance work must only be performed on the radiant heater by
authorised personnel. Only components that NUFER MEDICAL has expressly
approved for use with the CERATHERM 600-3 may be used.
After the unit is decommissioned, it must be disposed of correctly in
accordance with EU Directive 2002/96/EC. Get in touch with your allocated
NUFER MEDICAL sales partner.