13
Instruction Manual CERATHERM 600-3
4.

General Description

The built-in ceramic heater has very good radiation properties and generates invisible
infrared radiation of around 3 µm. This radiation spectrum is absorbed very effectively by the
skin and causes no harm to the patient's eyes.
5.

Assembly

5.1

Minimum Clearance

After you have removed the packaging, compare the
unit data on the specification plate with the electrical
connection data. Electricity is supplied via a 220–
240 V AC 50/60 Hz mains socket with a connection
value of 6 A.
Ensure that the wall- and ceiling mount (fixed or
mobile) is fastened into solid masonry (brickwork,
sand-lime brick or concrete) and that suitable wall-
plug screws are used.
Mount the heater head:
• with a clearance of at least 80 cm and no more
than 100 cm between the reclining surface and
the bottom edge of the heater head.
• with a clearance of at least 50 cm between the
ceiling/false ceiling and the top edge of the
heater head.
Ensure that there is always a clearance of at least 80 cm between the
reclining surface and the bottom edge of the heater head.
Ensure that there is always a clearance of at least 50 cm between the
ceiling/false ceiling and the top edge of the heater head.
The thickness of the underlay must also be taken into account.
Neglecting to do so may result in burns or fire.
14.171.C_GA_Ceratherm_600_3_EN.docx
Nufer Medical AG / 04.07.2014 / DU
Min.50 cm.
Figure 1
Seite 13 / 32