Nufer Medical CERATHERM 600-3 Instruction Manual

Hide thumbs Also See for CERATHERM 600-3:

Advertisement

CERATHERM 600-3
FOR THE OPERATORS AND THOSE RESPONSIBLE FOR THE CARE
• Read through this instruction manual thoroughly before using the
unit.
• Keep this instruction manual in a place where it can be referred to
whenever needed.
14.171.C_GA_Ceratherm_600_3_EN.docx
Radiant Heater
Instruction Manual
AND MAINTENANCE OF THIS UNIT:

Advertisement

Table of Contents
loading

Summary of Contents for Nufer Medical CERATHERM 600-3

  • Page 1 CERATHERM 600-3 Radiant Heater Instruction Manual FOR THE OPERATORS AND THOSE RESPONSIBLE FOR THE CARE AND MAINTENANCE OF THIS UNIT: • Read through this instruction manual thoroughly before using the unit. • Keep this instruction manual in a place where it can be referred to whenever needed.
  • Page 2: Table Of Contents

    ..................... 24 ENERAL INFORMATION 13. WARRANTY ......................... 25 13.1 ....................25 ARRANTY ERIOD 13.2 ..................25 ARRANTY ESTRICTIONS 14. DISPOSING OF THE UNIT ................... 25 15. SPECIFICATIONS* ....................... 26 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 2 / 32...
  • Page 3 )......29 EILING MOUNT WITH DOUBLE JOINT ARM LONG WITH HEIGHT ADJUSTMENT OPTIONAL 16.5 ) ................29 OBILE STAND WITH HEIGHT ADJUSTMENT OPTIONAL 17. REPAIR REQUEST FORM ................... 31 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 3 / 32...
  • Page 4: Introduction

    NUFER MEDICAL is in no way responsible for errors or accidents that are the result of non- compliance with the instructions on operation, care, maintenance and any other content given in this instruction manual.
  • Page 5: Scope Of Delivery

    NUFER MEDICAL or the local trained sales partner. If repair work is required, the unit must be returned to the NUFER MEDICAL sales partner located in your country, along with a completed ‘Repair Request Form’ (see chapter 17). Further...
  • Page 6: Definitions And Symbols

    This lists includes all the symbols that are displayed on the unit to aid operation or as a warning or note: Symbol Explanation Ensure that the instruction manual is observed. 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 6 / 32...
  • Page 7 Warning sign: Do not reach into the heater head -> Risk of burns. Warning sign: Do not place any objects on the protective grille -> Risk of fire. 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 7 / 32...
  • Page 8 If the unit is used in a room that corresponds to Category 3 or 4 under Swiss Low-Voltage Installation Standard NIN / 7.10, this connection must be connected to the room’s potential equalisation connection. 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 8 / 32...
  • Page 9: Safety Guidelines

    Monitoring is provided by TÜV SÜD Product Service GmbH, 80339 0123 München. The unit must be disposed of in accordance with EU Directive 2002/96/EC. Get in touch with your allocated NUFER MEDICAL sales partner. Safety guidelines Please ensure that the following safety guidelines are observed. The guidelines must be...
  • Page 10 90%. Ensure that no condensation forms on the unit. External conditions (such as temperature and drafts) and other equipment in the vicinity (e.g. phototherapy equipment) can affect the patient’s body temperature. 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 10 / 32...
  • Page 11 The patient must never be left unattended beneath the radiant heater while it is switched on -> Risk to patient safety. It is vital that care staff independently monitor the patient’s temperature -> Risk to patient safety. 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 11 / 32...
  • Page 12 3 (75%). The radiant heater must only be operated with original accessories that NUFER MEDICAL has expressly approved for use with the CERATHERM 600- 3. Observe the relevant information in chapter 16. In the event of a technical alarm shut the unit down immediately and have it checked by an authorised person and repaired as required.
  • Page 13: General Description

    The thickness of the underlay must also be taken into account. Neglecting to do so may result in burns or fire. 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 13 / 32...
  • Page 14: Product Description

    Instruction Manual CERATHERM 600-3 Product description The CERATHERM 600-3 radiant heater has four programmable heat output levels, which can be selected directly by pressing the buttons [heat level 1-4]. The default settings for the heat levels are 20, 50, 75 and 98%. All levels can be adjusted by a specialist upon request (within certain limits, e.g.
  • Page 15: Description Of Parts

    Radiates heat Hand grip Adjusts the radiant heater LED light Lights the patient -> Observe the relevant information in chapter 15! Ceramic heating element Generates heat Table 1 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 15 / 32...
  • Page 16 Main fuses 2 × 3.15 AT 20 Potential equalisation connection Connects to room’s potential connection (room categories 3 and 4) 21 Protective grille Functions as heat guard and provides ventilation Table 2 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 16 / 32...
  • Page 17: Operation

    Instruction Manual CERATHERM 600-3 Operation Operation of the CERATHERM 600-3 radiant heater has been kept deliberately straightforward. Use of the unit is largely self-explanatory. Master switch The unit’s ON/OFF switch is located on the rear side. This is normally switched on at all times.
  • Page 18: Control Elements On The Front

    When the buttons are referred to in the text below, the number will be Note given in square brackets [ ]. Given output corresponds to default settings. Cannot be any higher than 98% because of the internal hardware monitoring circuit. 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 18 / 32...
  • Page 19: Commissioning

    6. Once the controls have been set to standby mode via on/off button [7] on the front, the unit can be fully shut down by pressing mains switch [1] on the rear side. Adjusting the Heat Output [10] CERATHERM 600-3 Figure 6 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 19 / 32...
  • Page 20: Practical Examples Of Output Level Settings

    20% for safety reasons. The red alarm LED displays this status visually until the heater head is either switched off with on/off button [7] or the alarm is confirmed with 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 20 / 32...
  • Page 21: Led Lighting

    The supportive alarm monitoring function is not activated at safety level. LED Lighting The CERATHERM 600-3 is fitted with a modern LED lamp. This cutting-edge light source has a lower energy consumption and higher service life than other lights. -> Observe the...
  • Page 22: Troubleshooting

    Instruction Manual CERATHERM 600-3 Troubleshooting The CERATHERM 600-3 controls are able to detect and report the following malfunctions: 1. Supply voltage interruption (mains failure) 2. Circuit breaker interruption (triac) 3. Circuit breaker short-circuit (triac) 4. Safety switch interruption (relay) 5. Safety switch short-circuit (relay) 6.
  • Page 23: Testing The Control Elements

    Table 5 Testing the Control Elements When the controls are switched on, all relevant components (illuminated buttons, buzzer) are briefly switched on to check whether they are functioning correctly. 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 23 / 32...
  • Page 24: Cleaning/Disinfection

    The radiant heater must not be autoclaved or immersed in liquid. The radiant heater must only be disinfected by wiping over surfaces. Only alcoholic disinfectants may be used. 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 24 / 32...
  • Page 25: Warranty

    Warranty 13.1 Warranty Period The warranty period for CERATHERM 600-3 is given in the documents enclosed with the unit. Different warranty periods may apply for different countries. The warranty covers the repair of the unit and the replacement of defective parts in the event of manufacturing or material faults.
  • Page 26: Specifications

    Every 12 months as per service manual specifications, chapter 20 In accordance with Directive 2002/96/EC. Get in touch with your Disposal allocated sales partner. Dimensions 270 × 105 × 540 mm (W × H × L) 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 26 / 32...
  • Page 27: Accessories

    CNS (stainless steel) Material, reflector CNS (stainless steel) *NUFER MEDICAL reserves the right to make changes to the unit that may affect these specifications, at any time and without notice. Accessories Regarding the accessories approved for the CERATHERM 600-3 radiant heater: With regard to assembly and use, observe the assembly and operating instructions supplied with the accessories.
  • Page 28: Wall Mount With Double-Joint Arm (Optional)

    Pivotable, 360°. No height adjustment. Arm length as per drawing. Article number: 521A-60022-3 16.3 Wall mount with double-joint arm, long, with height adjustment (optional) Pivotable, 360°. With height adjustment. Arm length as per drawing. Article number: 521A-60025-3 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 28 / 32...
  • Page 29: Ceiling Mount With Double-Joint Arm, Long, With Height Adjustment (Optional)

    Article number: 521A-60024-3 Ceiling tube length customer-specific. 16.5 Mobile stand, with height adjustment (optional) With height adjustment. Triple-bar trolley base with two brakes. Article number: 521A-60035-3 Dimensions as per drawing. 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 29 / 32...
  • Page 30 Instruction Manual CERATHERM 600-3 Notes: 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 30 / 32...
  • Page 31: Repair Request Form

    Tel. : Invoice number: Model: CERATHERM 600-3 Accessories: Serial number: Detailed description of failure or problem: Anticipated work/repair: Repair Warranty repair Delivery of replacement unit Other Description: Date: Signature: 14.171.C_GA_Ceratherm_600_3_EN.docx Nufer Medical AG / 04.07.2014 / DU Seite 31 / 32...
  • Page 32 Valid Status 04/07/2014 Date Doc. version Version V1.2 or later Software version Varitronic AG / A.H., Nufer Medical AG / DU Authors 14.171.C_GA_Ceratherm_600_3_EN File name NUFER MEDICAL AG Morgenstrasse 148 CH-3018 Bern / Switzerland Tel. +41 (0)31 958 66 66 Fax +41 (0)31 951 46 73 info@nufer-medical.ch...

Table of Contents