Press the CLEANING Keypad Button to change the PROD. UNIT from MΩ.cm to μS/cm.
S E R V I C E : O P T I O N S
P R O D . U N I T
M Ω . c m
Press the MENU Keypad Button to return to PRE OPERATE Mode.
S E R V I C E : O P T I O N S
P R O D . U N I T
μ S / c m
When the Milli-Q is put into PRODUCT Mode, the LCD will look like :
Note that a Conductivity Measurement of 0.055 μS/cm is the same as a Resistivity measurement of
18.2 MΩ.cm.
You have finished changing to units of Conductivity.
Milli-Q Synthesis/Synthesis A10
OPERATE / STANDBY
OPERATE / STANDBY
P R E
O P E R A T E
POWER
SERVICE
ALARM
MEASURE
CLEANING
MENU
Press CLEANING
POWER
SERVICE
ALARM
MEASURE
CLEANING
MENU
Press MENU for
about 2 seconds
0 . 0 5 5 μ S / c m @ 2 5 ° C
T O C :
I
NSTALLATION
S E R V I C E : O P T I O N S
P R O D . U N I T
μ S / c m
P R E
O P E R A T E
4
p p b