Input Power - GE LOGIQ V5 Service Manual

Hide thumbs Also See for LOGIQ V5:
Table of Contents

Advertisement

GE
D
5496012-100, R
IRECTION
10-4-3

Input Power

10-4-3-1
Mains Cable Inspection
Table 10-6
Step
1
Unplug Cord Disconnect the mains cable from the wall and system.
2
3
4
10-4-4
Cleaning
10-4-4-1
General Cleaning
Table 10-7
Step
1
Console
2
Probe Holder Clean probe holders (they may need to be soaked to remove excess gel).
NOTE:
For your convenience or of the air filter is too dirty, replacement filters are available. refer to
Chapter 9 for the air filter replacement part number.
10-4-5
Physical Inspection
Table 10-8
Step
1
2
Scratches & Dents Inspect the console for dents, scratches or cracks.
Wheels & Brakes Check all wheels and casters for wear and verify operation of foot brake, to stop the unit from moving,
3
4
5
6
7
Op Panel Lights
8
Monitor Light
9
3
EVISION
Mains Cable Inspection
Item
Inspect
Inspect it and its connectors for damage of any kind.
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that no
Verify
strands may cause a short circuit.
Verify
Inlet connector retainer is functional.
General Cleaning
Item
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. Be careful
not to get the cloth too wet so that moisture does not enter the console.
Physical Checks
Item
Verify that all system labeling is present and in readable condition. refer to the LOGIQ V5/LOGIQ V3
Labeling
User Manual for details.
and release mechanism. Check all caster locks and caster swivel locks for proper operation.
Cables &
Check all internal cable harnesses and connectors for wear and secure connector seating. Pay
Connectors
special attention to footswitch assembly and probe strain or bend reliefs.
Shielding &
Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Covers
Missing covers and hardware could cause EMI/RFI problems while scanning.
External I/O
Check all connectors for damage and verify that the labeling is good.
Check for proper operation of all operator panel and TGC lights.
Check for proper operation of any monitor lights if available.
External
Check for proper operation of any external microphones by recording an audio test.
Microphone
Section 10-4 - Tools Required
B
Description
Description
Description
LOGIQ V5/LOGIQ V3
S
M
ASIC
ERVICE
ANUAL
10 - 7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Logiq v3

Table of Contents