GE
D
5496012-100, R
IRECTION
Section 10-3
Maintenance Task Schedule
Table 10-2 Customer Care Schedule
Service at Indicated Time
Clean Probe Holders
Clean Air Filter
Inspect AC Mains Cable
Inspect Cables and Connectors
Clean Console
Inspect Wheels, Casters, brakes and
Swivel Locks
Check Control Panel Movement
Console Leakage Current Checks
Peripheral Leakage Current Checks
Surface Probe Leakage Current Checks
Endocavity Probe Leakage
Current Checks
Transesphongeal Probe Leakage
Current Checks
Surgical Probe Leakage
Current Checks
Measurement Accuracy Checks
Functional Checks
NOTE:
PMs are not mandatory, the table above is for reference only.
NOTE:
May require specialized equipment to complete.
3
EVISION
(cont'd)
Daily
Weekly Monthly
•
•
•
•
•
•
•
-
LOGIQ V5/LOGIQ V3
B
S
ASIC
Per Facilities
QA Program
Notes
more frequently depending on
your environment
Mobile Unit Check Weekly
Mobile Unit Check Daily
Mobile Unit Check Daily
also after corrective
•
maintenance
also after corrective
•
maintenance
also after corrective
•
maintenance
also after corrective
•
maintenance
also after corrective
•
maintenance
also after corrective
•
maintenance
also after corrective
•
maintenance
also after corrective
•
maintenance
M
ERVICE
ANUAL
10 - 3