1.
From the power measurement screen, select
"Laser" and enter.
2.
With the up/down navigation keys, select the first
wavelength desired. Press the right navigation key
to enter the change wavelength screen.
3.
Enter the wavelength value by pressing the
up/down navigation keys to change each number
and the right/left navigation keys to move to the
next number. After you are finished, press "Save"
and enter.
4.
Repeat steps 2 and 3 for other wavelengths desired.
When finished, press Enter. Those values for which
"Save" was pressed will be default startup values,
the others will be saved until the instrument is
turned off.

5.3 Startup Defaults

If you are planning to do similar measurements most of
the time, you can set up the Vega so that it automatically
goes into the measurement mode you want upon startup.
Since these desired settings are saved in the sensor
"smart connector", you can save different settings for
different sensors. The following section describes how to
set up the Vega with the PD300 sensor so that it will start
up in the desired configuration.
Setting up the startup defaults: (See section 5.4 for
further details)
1. With the sensor connected, turn on the Vega. Select
"Menu" and enter. Select "Configure" and enter.
2. Select "Filter" and enter. Select IN or OUT as the
startup default and enter
See Section 5.4.3.
3. Select "Display Mode" and select the desired
startup screen. Press "Save". (Note that this
selection is saved in the Vega and not the sensor so
will be operative for all sensors). For more details on
the screen types see section 4.6.
4. Select "Power Range:" and in the same way choose
the manual power range you wish to be the default or
.
For details of filter in/out
61