Screening Alarm; General; Procedure - FLIR T4 series User Manual

Hide thumbs Also See for FLIR T4 series:
Table of Contents

Advertisement

22

Screening alarm

22.1 General

The screening alarm can be used, for example, at airports to detect passengers with ele-
vated body temperatures, which may indicate the presence of a fever.
Activating the screening alarm will turn on a measurement box and screening data in the
result table.
The sampled average temperature.
The alarm temperature.
The measured temperature.
The alarm will trigger when the measurement box measures a temperature higher than
the alarm temperature. The alarm temperature is, in turn, the sum of a specified allowed
deviation and a sampled average value.

22.2 Procedure

Follow this procedure:
1. Enable the screening mode by selecting
camera > Screening mode = On.
2. Push the joystick to display the menu system.
3. Use the joystick to go to
4. Push the joystick. This displays a submenu.
5. Select
(Screening).
6. Push the joystick. This displays a dialog box where you can define the settings for the
alarm.
• Allowed deviation: The allowed deviation from the sampled average.
• Alarm sound: Applicable values are Beep or No sound.
7. Push the joystick. This closes the dialog box.
8. Aim the camera toward a point of interest. The object should be within the frame of
the measurement box.
9. Push and hold the programmable button
10. Push the programmable button
11. Aim the camera toward more points of interest. Sample 10 times to build up a sample
base by pushing the programmable button
The alarm is now set up and ready to use. Occasionally record a few samples if the
alarm is used for a long time or if the conditions change.
Note
• The algorithm has a memory of the last 10 samples. It discriminates between the high-
est and lowest values, and calculates an average of the remaining values.
• Do not modify the measurement setup or activate another alarm because this will de-
activate the screening alarm.
#T559879; r. AP/36066/36066; en-US
(Settings) > Device settings > Set up
(Recording mode).
to reset the sampled average.
to sample.
.