Functions; Vac Function; Vdc Function; Low Ohm And Continuity Function - AEMC 1026 User Manual

Digital/analog megohmmeter
Table of Contents

Advertisement

Functions

ACV Function:
Turn the function selector to the voltage range. Connect the black test lead
to (–) terminal and the red one to the (+) terminal. Connect the test lead to
the circuit in test. Reading can be held by pressing the HOLD key.
DCV Function:
Turn the function selector to the voltage range. Connect the black test lead
to (–) terminal and the red one to the (+) terminal. Press the TEST button
to change the mode from V
circuit in test. Reading can be held by pressing the HOLD key.
Low Ohm Function and Continuity Function:
Turn the function selector to the 4000 range. Connect the black test lead
to (–) terminal and the red one to the (+) terminal. Connect the test lead to
the circuit in test. If the reading is less than 40, the continuity beeper will
sound.
Because the test current provided by the meter could reach 200mA, do not
use this range to test electronic components like diodes, transistors or
fuses.

MegOhm Insulation Resistance Test Function:

Turn the function selector to desired test voltage range. The LCD displays
"----" to indicate the megohmmeter is standing by.
uality Products Online at:
Warning:
Do not test insulation on a live sample and make sure it is fully
discharged before each test. The Model 1026 is equipped with
V
function which may be used to check if the sample is live.
AC/DC
GlobalTestSupply
www.
Digital Megohmmeter Model 1026
to V
function. Connect the test lead to the
AC
DC
- 9 -
.com
sales@GlobalTestSupp