Aluminum, Ammonia & Sulfide - eXact Micro 20 Instruction Manual

Powerful dual-wavelength photometer system with bluetooth smart
Hide thumbs Also See for eXact Micro 20:
Table of Contents

Advertisement

Aluminum, Ammonia & Sulfide test procedure
1
TURN METER ON
Press the
the current selection. It will then display the last reading.
SELECT GROUP & MENU
2
Press and re–press the
select the test parameter (see chart on page 10–11).
RINSE & FILL CELL WITH SAMPLE
3
Rinse the CELL at least 3 times with the water sample you will be testing—rinsing minimizes
the potential for cross–contamination from a previous test. Finally, fill CELL to capacity with
the water sample. Tilt meter to discard about 0.2mL water to leave room for liquid reagent.
Follow the steps to 'SELECT CUSTOMER' & 'CONNECT DEVICE VIA BLUETOOTH' before
proceeding (see page 8).
ADD DROPS
4
Using the selected bottle of reagent add the required drops (see chart on page 10–11) and
cover the CELL with the CELL COVER. Precaution: Ensure that the bottle is straight when
dispensing drops.
ZERO METER*
5
Press the
indicate that the sample is ready for testing.
DIP STRIP & PRESS READ
6
Using the required strip (see chart on page 10–11), dip strip into the CELL, and immediately
press
to initiate a 20 second countdown. Move the strip using a gentle back and forth
motion (approx. 2 strokes/sec). Remove and discard the strip after "1" on the display
disappears*. The meter will automatically start to count up. The count up time will vary for
each parameter. At the end, the cursor will move across the display while the meter prepares
to measure the sample. Record the value displayed for the respective parameter This value
is automatically stored in its MENU and if using the eXact iDip
the app's 'HISTORY. After testing, rinse CELL immediately and clean with the brush. After Sulfide
testing: rinse CELL with Distilled White Vinegar, 0.1N HCl, or Muriatic Acid and clean with brush.
Aluminum, Ammonia & Sulfide (SPECIAL NOTES)
Aluminum — A. First, clean the CELL with 0.1N HCl, Distilled Vinegar (5%), or Muriatic Acid before
1
filling the meter with the sample to be tested. B. If running multiple tests in a row, using the same
water sample, the CELL does not have to be rinsed or cleaned with acid between each test. It is
recommended that the CELL be rinsed 3 times with the sample water.
Ammonia — The calibration of the meter is based on a water temperature between 14°C (57°F) and
2
28°C (82°F). If temperature is below 14°C (57°F), your final Ammonia value may read low.
Sulfide — A. For results as Hydrogen Sulfide (H2S), multiply the resulting value by 1.06. B. The
3
calibration of the meter is based on the water sample temperature above 20°C (68°F). If the water
sample is below 20°C (68°F), the strip has to dip in the sample for an additional 10 seconds.
20
button to power the meter on; the display will show annunciators, followed by
button to Select Group. Press and re–press the
button. The cursor will move across the display followed by 0.00 PPM. This will
*For best accuracy, when testing outdoors in sunlight, place cap over cell cover
app, the result will be saved in
®
button to