Pump Calibration Pass
If the pump calibration/pump test passes, the audible alarm
beeps three times, the display shows rECAL SuccESS, and
the detector returns to normal operation.
Pump Calibration Fail
If the pump calibration/pump test fails, the audible alarm
beeps six times and the display reads PumP FAiLurE.
If the pump calibration fails, try to calibrate it again. If the
calibration continues to fail, replace the pump. (Refer to
Replacing a Sensor, Pump, or Pump
GasAlertMax
Pump Calibration
Filter.)