Problem
The activation self-test fails during
one of the checks.
The detector does not display normal
ambient gas reading after the
activation self-test.
The detector does not respond to the
pushbuttons.
The detector does not accurately
measure gas.
Table 15. Troubleshooting Tips (cont.)
Possible Cause
General fault
Sensor not stabilized
Detector requires calibration
Target gas is present
Battery is depleted
Detector is performing operations
that do not require user input
Detector requires calibration
Detector is colder/hotter than
ambient gas
Sensor screen is blocked
GasAlertMax
Troubleshooting
Solution
Contact BW
Used sensor: wait 60 seconds
New sensor: wait 5 minutes
Calibrate detector
Detector is operating properly. Use
caution in suspect areas.
Replace battery
Pushbutton operation restored
automatically when the operation
ends
Calibrate detector
Allow detector to acquire ambient
temperature before use
Clean sensor screen
53