Problem
The activation self-test fails during
one of the checks.
The detector does not display normal
ambient gas reading after the
activation self-test.
The detector does not respond to the
pushbuttons.
The detector does not accurately
measure gas.
Table 15. Troubleshooting Tips (cont.)
Possible Cause
→
General fault
→
Sensor not stabilized
→
Detector requires calibration
→
Target gas is present
→
Battery is depleted
→
Detector is performing operations
that do not require user input
→
Detector requires calibration
→
Detector is colder/hotter than
ambient gas
→
Sensor screen is blocked
GasAlertMax
Troubleshooting
Solution
→
Contact BW
→
Used sensor: wait 60 seconds
New sensor: wait 5 minutes
→
Calibrate detector
→
Detector is operating properly. Use
caution in suspect areas.
→
Replace battery
→
Pushbutton operation restored
automatically when the operation
ends
→
Calibrate detector
→
Allow detector to acquire ambient
temperature before use
→
Clean sensor screen