Low Battery Alarm; Automatic Shutdown Alarm - BW Technologies GasAlertMax User Manual

H2s, co, o2, combustibles 2, 3, and 4 gas detectors
Hide thumbs Also See for GasAlertMax:
Table of Contents

Advertisement

Pump Failure Alarm
This alarm activates if the startup
pump test has failed or the pump is
missing/broken.
The pump may alarm when attaching a calibration
hose or replacing the filter.

Low Battery Alarm

The detector tests the battery on activation and continuously
thereafter. If the battery voltage is low, the detector activates
the low battery alarm.
The low battery alarm continues until you replace the battery
or the battery power is almost depleted. If the battery voltage
drops too low, the detector executes an automatic shutdown.
If the confidence beep is on, the audible alarm does
not beep during a low battery alarm. (See the
Confidence Beep
section.)
Note
Note

Automatic Shutdown Alarm

If the battery voltage is in immediate danger of dropping
below the minimum operating voltage, the audible alarm
beeps 8 times and the visual alarm flashes 8 times. After 3
seconds, the display blanks out and the detector stops
normal operation. The display shows I LOW periodically
until the battery power is depleted.
For directions on how to replace the battery, see the section
Replacing the
Battery.
Typically, the low battery alarm continues for 30
minutes before an automatic shutdown.
GasAlertMax
Alarms
Note
31