Stopping A Gas Alarm; Sensor Alarm; Pump Alarms - BW Technologies GasAlertMax User Manual

H2s, co, o2, combustibles 2, 3, and 4 gas detectors
Hide thumbs Also See for GasAlertMax:
Table of Contents

Advertisement

GasAlertMax
User Manual
To change the factory-set alarm setpoints, refer to the
Calibration and Setting Alarm Setpoints
You can disable an alarm by setting the alarm
setpoint to 0 (display: - -).

Stopping a Gas Alarm

The low, high, and STEL alarm stops when the ambient gas
level returns to the acceptable range.
The detector computes the TWA value based on an 8-hour
workday. Only deactivating the detector can stop the TWA
alarm.
Acknowledge Latched Alarm
If the latched alarm function is turned on, the audible and
visual alarms continue to sound and flash until the low, high,
or STEL alarm condition is acknowledged. Press C to
deactivate the audible and visual alarms when the current
ambient gas reading falls below the low alarm level. The
alarms cannot be deactivated if an alarm condition is still
present. A TWA alarm condition will not reset.
A STEL alarm may take up to 15 minutes of clean
air readings before the latched alarm can be
deactivated.
30
section.
Note
Note

Sensor Alarm

The detector tests for a missing or defective sensor during
the activation self-test. See the

Pump Alarms

The internal pump draws air over the sensors. If the pump
stops working or the pump filter becomes clogged, the
detector activates the pump alarm.
Pump Alarm
Turn off the detector immediately and
clear the pump inlet, hose, and filter.
Blocked Pump Alarm
If the pump alarm continues and the
LCD advises the inlet is blocked, see
the
Pump Calibration
section.
Troubleshooting
section.