Confidence Beep; User Options Menu - BW Technologies GasAlertMax User Manual

H2s, co, o2, combustibles 2, 3, and 4 gas detectors
Hide thumbs Also See for GasAlertMax:
Table of Contents

Advertisement

Confidence Beep

The confidence beep tells you the detector is on and the
battery has sufficient power to respond to a hazardous level
of gas and emit an alarm. Instead of beeping when the
battery's power is low, the audible alarm beeps to advise you
the battery has sufficient power. The confidence beep stops
when the battery power is low.
You can only activate the confidence beep at startup.
To turn on the confidence beep:
1.
If the detector is on, deactivate the detector.
Press F while pressing C at startup.
2.
After the self-test completes, the detector continuously beeps
once every 5 seconds.
To turn off the confidence beep, repeat steps 1 and 2 above.

User Options Menu

The following are the available user options:
1.
Finish options and exit the user options menu.
2.
Adjust real-time clock and calendar. (User
downloadable datalogger model only)
3.
Enable or disable operation of one or more
sensors.
4.
Enable or disable pass code protection.
5.
Set latching alarm function.
6.
Enable or disable the automatic Oxygen calibration
at startup.
7.
Set the calibration gas span concentration values.
8.
Enable or disable the startup pump test.
9.
Set the STEL calculation period.
10. Set the LCD to display in German.
GasAlertMax
Confidence Beep
15

Advertisement

Table of Contents
loading