RF Analyzer
N9912A
Before You Begin
Before checking, verifying, or adjusting the analyzer, refer to the following paragraphs to:
• Make sure the operating environment is within 19 to 27 °C and 5 to 95% humidity.
• Make sure that proper electrostatic discharge (ESD) protection is provided.
• Review the principles of connector care.
Protect Against Electrostatic Discharge (ESD)
This is important. If not properly protected against, electrostatic discharge can seriously damage your
analyzer, resulting in costly repair.
CAUTION
To reduce the chance of electrostatic discharge, follow all of the recommendations outlined
in
"Electrostatic Discharge Protection" on page
Allow the Analyzer to Warm Up
Before testing or adjusting the N9912A analyzer, allow it to warm up as described under "DUT Warmup" in
the "Getting Started" section of the N7841A help file.
Service Guide N9912-90003
Tests and Adjustments
1-5, for all of the procedures in this chapter.
Before You Begin
3-3