Problems And Solutions - SPHINX Myth3 Service Manual

Table of Contents

Advertisement

SPHINX Myth 3

8. PROBLEMS AND SOLUTIONS

At the moment of writing the Myth 3 has no known
specific problems.
If in the future you encounter any problem(s) you
may enter the info in this table. This table can then
be used for future reference.
Problem
Bandwidth ≈ 100 kHz
Cause
Value of C2L, C2R too large
(220 pF)
Service Manual
Please also send (by fax or e-mail) the specific
information to the Sphinx International Service
Department (see page 3): this info can then be
added to the general database to aid others.
Solution
Change C2L, C2R value to 47
pF
Refer to
page...
17/18
11