3
Wiring
3.1
Connect input and output wiring
Connect input and output wiring to the appropriate terminals on the input/output wiring
connector, which is the far right connector. Refer to
inserted into the sleeve on the top panel (shown in
•
Use 24 to 16 AWG (0,25 to 1,5 mm
•
Ground the shields at a single point only.
Table 3-1: Input/output wiring terminals
c 2+
c 4 +
c 6 +
c 8 +
c 10 +
c 12 +
c 14 +
c 16 +
c 18 +
c 32 (B line)
Figure 3-1: Input/output wiring terminal card
3.2
Connect the Model 3500 to the sensor
If you are installing the Model 3300 applications peripheral, this step is not required. Go to
Section
To connect the Model 3500 transmitter to a Micro Motion sensor, follow the following
instructions.
Installation Manual
Terminal number
a 2 –
a 4 –
a 6 –
a 8 –
a 10 –
a 12 –
a 14 –
a 16 –
a 18 –
a 32 (A line)
3.4.
Table 3-1
Figure
3-1).
2
) twisted-pair shielded wire.
Designation
Primary 4–20 mA output / HART
Secondary 4–20 mA output
Frequency input
Discrete input 1
Discrete input 2
Frequency output
Discrete output 1
Discrete output 2
Discrete output 3
RS-485 output
Wiring
and to the card that is
13