Quick Start Guide
5 

Connect the wiring and power up

Figure 5-1
Rosemount 3051 Transmitter and enable communications with a
communication device.
Figure 5-1: Transmitter wiring diagrams
A. Power supply
B. Ground
C. Resistor
Table 5-1: Power supply and resistance needs by communicator
type
Communicator
AMS Device Manager
AMS Trex (HART
AMS Trex (HART + pwr)
AMS Device
Configurator
Bluetooth
Quick Service buttons
Local operator interface
(LOI)
Resistance and power requirements vary based on the interface type
used to communicate with the device. See
power and resistance needs.
16
shows wiring connections necessary to power a
Power supply
≥ 16.6 Vdc
)
≥ 16.6 Vdc
®
None
≥ 10.5 Vdc
App
®
≥ 10.5 Vdc
≥ 10.5 Vdc
December 2023
Resistor
≥ 250 Ω
≥ 250 Ω
None
None
None
None
Table 5-1
for specific loop