problem
Different
result in
comparison
with another
meter.
The meter
does not
power on.
Test does not
Insufficient sample or there
start after
is a problem with test strip
applying
or meter.
sample.
The result
is doubtful.
caused by
Meter is calibrated against
whole blood from the manu-
facturer.
Blood samples are not taken
at the same time.
Battery is flat or there is a
problem with the meter.
There may be a problem
with the test strip.
The code numbers of the
meter and of the test strip
vial do not match.
what to do?
To check the accuracy of your
system use Omnitest® 3
Control solution or measure
against a laboratory device
which uses plasma.
Repeat the measurement and
take the samples at the same
time for comparison.
Change batteries and if the
problem persists, contact your
distributor or nearest B. Braun
representative.
Apply sufficient amount of a
fresh sample to a new blood
glucose test strip and re-test.
Conduct a control solution
test or a check strip test.
Re-test with a new test strip.
Conduct a control solution
test.
Re-test with a new test strip.
If the problem persists, please
contact your authorized repre-
sentative or customer support.