OPTIFLEX 1300 C

6.4.12 How to decrease the length of probes

INFORMATION!
This information is for:
Ø4 mm / 0.15
single rod,
Ø2 mm / 0.08
Ø4 mm / 0.15
WARNING!
Do not change the length of coaxial, Ø8 mm / 0.3
damage the probe.
How to decrease the length of single rod probes
• Cut the rod to the correct length.
• Measure the length of the rod from the flange facing.
• Go to Supervisor > Advanced Setup > Installation Setup > Probe Length
Supervisor > Advanced Setup > Installation Setup > Probe Length.
Supervisor > Advanced Setup > Installation Setup > Probe Length
Supervisor > Advanced Setup > Installation Setup > Probe Length
• Type in the new value. Press ^.
• Press > and
• Select Save
Save and press ^.
Save
Save
INFORMATION!
If you cannot measure the length of the probe, you can use Quick Setup > Setup Mode > Auto
Probe Length
Probe Length as an alternative. But you must do this procedure when the tank is empty and the
Probe Length
Probe Length
device is installed on the tank.
How to decrease the length of cable probes
• Cut the cable to the correct length.
• Attach the cable to the counterweight. Tighten the socket set screws.
Use a 3 mm Allen wrench.
i
• Loosen the socket set screws that hold the counterweight.
Use a 3 mm Allen wrench.
i
• Remove the counterweight.
• Measure the length of the cable from the flange facing to the bottom of the counterweight.
Add the length of the counterweight to get the total probe length. Subtract the length of the
i
cable engaged in the counterweight. Refer to the diagram after this procedure.
• Go to Supervisor > Advanced Setup > Installation Setup > Probe Length
Supervisor > Advanced Setup > Installation Setup > Probe Length.
Supervisor > Advanced Setup > Installation Setup > Probe Length
Supervisor > Advanced Setup > Installation Setup > Probe Length
• Type in the new value. Press ^.
• Press > and
• Select Save
Save and press ^.
Save
Save
INFORMATION!
If you cannot measure the length of the probe, you can use Quick Setup > Setup Mode > Auto
Probe Length as an alternative. But you must do this procedure when the tank is empty and the
Probe Length
Probe Length
Probe Length
device is installed on top of the tank.
09/2010 - 4000172305 - HB OPTIFLEX 1300 R06 en
double cable,
¨
single cable and
¨
single cable probes only
¨
(Esc) at the same time to exit to the "save settings" window.
(Esc) at the same time to exit to the "save settings" window.
www.krohne.com
single cable and double rod probes. This will
¨
Quick Setup > Setup Mode > Auto
Quick Setup > Setup Mode > Auto
Quick Setup > Setup Mode > Auto
Quick Setup > Setup Mode > Auto
Quick Setup > Setup Mode > Auto
Quick Setup > Setup Mode > Auto