Error Handling - KROHNE OPTIFLEX 1300 C Handbook

Level meter
Hide thumbs Also See for OPTIFLEX 1300 C:
Table of Contents

Advertisement

OPTIFLEX 1300 C
How to get more data about the error (error records function)
• Select an error log and press ^ ^ ^ ^ to read the help text.
i
Typical data is given in the illustration that follows.
• For solutions to the problem, refer to "Error handling" section.
Figure 6-16: Description of the error
1 Description of the error
2 Time since error occurred in Days:Hours:Minutes:Seconds
3 Type of error
The time since the error occurred is measured in Days:Hours:Minutes:Seconds
the time when the device is energized. The error is saved in the memory of the device when it is
switched off. The counter continues when the device is switched back on.
The error symbol is no longer displayed after you read the error record.

6.6.2 Error handling

Types of error message
Type of error
Error
Warning
09/2010 - 4000172305 - HB OPTIFLEX 1300 R06 en
Days:Hours:Minutes:Seconds. It only includes
Days:Hours:Minutes:Seconds
Days:Hours:Minutes:Seconds
Error
code
E
If an error message is displayed in the Error Records (menu
item B.2.12), the measured value is not correct. The output
current goes to the value set in menu items C.3.5 Error
Handling (Output 1) and C.4.5 Error Handling (Output 2). For
more data about the menu items, refer to
on page 82.
W
A condition that may not have an immediate effect on device
performance.
www.krohne.com
OPERATION
Description
C. Advanced setup
6
107

Advertisement

Table of Contents
loading

Table of Contents