Page 2
PROFIBUS-DP und das PROFIBUS-Logo sind eingetragene Warenzeichen der PROFIBUS Nutzerorganisation e.V. (PNO) SIMATIC ist ein eingetragenes Warenzeichen der SIEMENS AG TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 2 of 139 TR - ELE - BA - DGB - 0022 - 04...
5.1 PROFIBUS-DP – Schnittstelle ....................22 5.1.1 Kommunikationsprotokoll DP .................. 22 5.2 Analog – Schnittstelle ......................23 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Page 4
8.5.12 Analogstrom im Fehlerfall ..................53 8.6 Konfigurationsbeispiel, SIMATIC Manager V5.1 ..............54 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 4 of 139 TR - ELE - BA - DGB - 0022 - 04...
Page 5
10.3 Laser-Brille ..........................68 10.4 Steckerabdeckung IP-65 ...................... 68 10.5 PROFIBUS / Versorgung - Gegenstecker ................69 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
– Neues Design 25.02.15 – Laser Lebensdauer Verweis auf Support-DVD entfernt 08.02.16 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 6 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Es gelten somit zusammen folgende Dokumentationen: anlagenspezifische Betriebsanleitungen des Betreibers, dieses Benutzerhandbuch TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 7 of 139...
Allgemeines 1.2 EG-Konformitätserklärung Die Mess-Systeme wurden unter Beachtung geltender europäischer bzw. internationaler Normen und Richtlinien entwickelt, konstruiert und gefertigt. Eine entsprechende Konformitätserklärung kann bei der Firma TR-Electronic GmbH angefordert werden. Hersteller Produkte, TR-Electronic GmbH D-78647 Trossingen, besitzt ein zertifiziertes Qualitätssicherungssystem gemäß ISO 9001.
Anlage/Maschine in die das Mess-System eingebaut werden soll, den Bestimmungen der EG-EMV-Richtlinie, den harmonisierten Normen, Europanormen oder den entsprechenden nationalen Normen entspricht. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016...
das Beachten der Betriebsanleitung des Maschinen- bzw. Anlagen-Herstellers, das Betreiben des Mess-Systems innerhalb der in den technischen Daten angegebenen Grenzwerten (Montageanleitung/Benutzerhandbuch). TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 10 of 139...
Eigenmächtige durchgeführte Reparaturen. Katastrophenfälle durch Fremdeinwirkung und höhere Gewalt. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 11 of 139...
Installation, Inbetriebnahme und Bedienung festlegen. Beaufsichtigungspflicht bei zu schulendem oder anzulernendem Personal ! TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 12 of 139 TR - ELE - BA - DGB - 0022 - 04...
DIN EN 60825-1 in der neuesten Fassung beachten. Es sind die geltenden gesetzlichen und örtlichen Bestimmungen zum Betrieb von Laseranlagen zu beachten. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Page 14
Linsenöffnung des Mess-Systems regelmäßig mit einem weichen Tuch reinigen. Reinigung keine aggressiven Reinigungsmittel Verdünner oder Aceton verwenden! TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 14 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Distanz bestimmt. Abbildung 1: Standard Anwendung Gerätedaten: Messbereich bei LLB-65/LLB-500 auf natürliche Oberflächen: 0,05 m bis ca. 65 m Messbereich bei LLB-500 auf reflektierende Zieltafel: 0,5 m bis ca. 500 m PROFIBUS-DP Schnittstelle Flexible Spannungsversorgung (13...30VDC) ...
Empfängeroptik Produkt Bezeichnungslabel, siehe Kapitel 2.9 auf Seite 13 3x M12 Stecker/Buchse, PROFIBUS-DP – Schnittstelle, Versorgung TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 16 of 139 TR - ELE - BA - DGB - 0022 - 04...
Distanzen (z.B. > 150 m beim LLB-500) relevant sein können. Diese Effekte sind beschrieben in: B.Edlen: “The Refractive Index of Air, Metrologia 2”, 71-80 (1966) TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016...
Fehlerhafte Messungen können auch dadurch entstehen, wenn der Laserstrahl von anderen Objekten entlang der Messstrecke reflektiert wird. Vermeiden sie reflektierende Objekte entlang der Messstrecke. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 18 of 139...
● 953 mbar Luftdruck Diese Effekte sind beschrieben in: B.Edlen: “The Refractive Index of Air, Metrologia 2”, 71-80 (1966) TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
1 digitaler Ausgang zur Fehleranzeige siehe 4.1 Messgenauigkeit auf Seite 19. Bei Dauermessbetrieb ist die max. Temperatur auf 45°C reduziert. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 20 of 139 TR - ELE - BA - DGB - 0022 - 04...
4.3 Geräteabmessungen Alle Abmessungen in mm TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 21 of 139...
Datenaustausch in der Feldebene konzipiert ist. Die Grundfunktionalität wird durch die Leistungsstufe V0 festgelegt. Dazu gehören der zyklische Datenaustausch sowie die stations-, modul- und kanalspezifische Diagnose. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 22 of 139...
Die Analog-Schnittstelle des LLB ist als Stromquelle (0…20mA oder 4…20mA) ausgelegt. Es können Lasten bis maximal 500 getrieben werden. Die Genauigkeit des analogen Ausgangs beträgt beim LLB-65 = ±0,2% und beim LLB-500 = ±0,1% auf den Messbereich. Conf Conf ...
Versorgungsspannung für das LLB zu benutzen. Für die Versorgungsspannung wird ein Kabelquerschnitt von min. 0,75 mm empfohlen. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 24 of 139 TR - ELE - BA - DGB - 0022 - 04...
Konfiguration siehe Seite 50 und 51. DOUT Abbildung 3: Beschaltung der Digital Ausgänge TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
EMV-Richtlinie sowie Schirmungs- Erdungsrichtlinien in den jeweils gültigen Fassungen zu beachten! TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 26 of 139 TR - ELE - BA - DGB - 0022 - 04...
Terminierungsschalter = ON abzu- schließen. In diesem Zustand wird der weiterführende PROFIBUS abgekoppelt. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Sicherstellen, dass der Gesamtwiderstand am Analogausgang kleiner als 500 ist. Konfiguration siehe Seite 49, 52 und 53. Abbildung 4: Verbindung eines analogen Anzeigeinstrumentes und einer SPS Steuerung TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 28 of 139...
Analog Masse Analog Ausgang 0…20 mA bzw. 4…20 mA Geräte Masse Geräte Masse TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 29 of 139...
(GND) und den Massepunkt für den Analogausgang (AGND). GND und AGND sind über ein RC-Glied mit dem Gehäuse verbunden, siehe Abbildung 5. Abbildung 5: Verbindung zwischen Abschirmung, Masse (GND) und Analog-Masse (AGND) TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 30 of 139...
POWER (grün): ON, wenn sich die Spannungsversorgung im zulässigen Bereich befindet Abbildung 6: Geräte-Status LEDs TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Bus angeschlossen wurden. Dadurch wird eine hohe Sicherheit gegenüber Projektierungsfehlern erreicht. Das Mess-System hat die Identnummer 0D65 (Hex). TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 32 of 139 TR - ELE - BA - DGB - 0022 - 04...
Inputs zurückgeben WCFG = Wait Configuration DXCHG = Data Exchange Abbildung 8: DP-Slave Initialisierung TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Mess-System Störung Entsprechende Maßnahmen im Fehlerfall siehe Kapitel „Diagnose über optische Bus- Status-Anzeigen“, Seite 59. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 34 of 139 TR - ELE - BA - DGB - 0022 - 04...
0 = Konsistenz über ein Byte oder Wort Konsistenz: 1 = Konsistenz über das ganze Modul TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
- Min Analogstrom - Analogstrom im Fehlerfall aus Sicht des Bus-Masters TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 36 of 139 TR - ELE - BA - DGB - 0022 - 04...
5 – 4 3 – 0 Data Konsistenz Wort Format Eingangsdaten Längen-Code TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 37 of 139...
Page 38
– 2 Data Default (dez.) 0 (–19685040 - +19685040) Offset TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 38 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Page 39
– 2 Data Default (dez.) 0 (–19685040 - +19685040) Fehlerwert TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 39 of 139...
5 – 4 3 – 0 Data Konsistenz Wort Format Eingangsdaten Längen-Code TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 40 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Page 41
1/10 mm (Default) MIL (1/1000 Inch) 1/100 Inch 1/10 Inch Inch TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 41 of 139...
Page 42
Wert eingetragener Fehlerwert Min Analogstrom 0 mA 4 mA TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 42 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Page 43
Data Default (dez.) 0 (–19685040 - +19685040) Digital Ausgang 1 On TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 43 of 139...
Page 44
Data Default (dez.) 0 (–19685040 - +19685040) Digital Ausgang 2 Off TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 44 of 139 TR - ELE - BA - DGB - 0022 - 04...
Page 45
Data 0 (0 – 200) Default (dez.) Analogstrom im Fehlerfall TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 45 of 139...
Es erfolgt keine Quittierung des Vorgangs über die Eingänge. Untergrenze Obergrenze entsprechend der eingestellten Auflösung, bezogen auf 500 m TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 46 of 139 TR - ELE - BA - DGB - 0022 - 04...
Obergrenze 0x01 2C 5E B0 (+19 685 040) 0x00 00 00 00 Default TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 47 of 139...
Mit dem Messzyklus wird die Abtastrate der Distanzmessungen eingestellt. Der Wert 0 bewirkt eine schnellstmöglichste Messung (Dauermessbetrieb). Diese kann, abhängig von der Oberfläche beim LLB-65 0,15 bis ca. 4 s und beim LLB-500 0,04 bis ca. 4 s andauern. Der Wert 0xFF deaktiviert diese Funktion und die Laser-Diode wird abgeschaltet.
Strombereich: 0…20 mA 0 mA Strombereich: 4…20 mA 4 mA TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 49 of 139...
Der Parameter „Messzyklus“ muss dazu auf dezimal „255“ stehen. Distanzmessung erfolgt automatisch so schnell wie möglich. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 50 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Obergrenze 0x01 2C 5E B0 (+19 685 040) Default 0x00 00 00 00 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 51 of 139...
Offsets bzw. Presets. Der Arbeitsbereich beginnt an der Messreferenz, siehe auch Seite 21, und endet bei dem LLB-65 bei 65 m und bei dem LLB-500 bei 500 m. Abweichungen hiervon ergeben eine fehlerhafte Stromausgabe! TR-Electronic GmbH 2011, All Rights Reserved...
Seite 40 Untergrenze 0x00 Obergrenze 0xC8 Default 0x00 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 53 of 139...
Dateinamen und Einträge in den nachfolgenden Masken sind nur als Beispiele für die Vorgehensweise zu betrachten. Zur Aufnahme der GSD-Datei in den Katalog, muss diese zuerst installiert werden: TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 54 of 139...
Page 55
Seite 40 Der Eintrag Universalmodul wird irrtümlicherweise automatisch von manchen Systemen bereitgestellt, darf jedoch nicht verwendet werden! TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Page 56
Netzeinstellungen vorgenommen werden (Klick mit rechter Maustaste auf das Mess- System-Symbol --> Objekteigenschaften): TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 56 of 139 TR - ELE - BA - DGB - 0022 - 04...
Page 57
Gewünschte Konfiguration aus dem Katalog auf den Steckplatz übertragen (Drag&Drop). Das Mess-System-Symbol muss aktiv sein. Parametrierung vornehmen mit Doppelklick auf die Steckplatznummer: TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
– Hardwarefehler – Mess-System tauschen Siehe auch Kapitel „Geräte-Statusanzeige“ auf Seite 31. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 58 of 139 TR - ELE - BA - DGB - 0022 - 04...
- Analog Ausgang Min und Max – Gerät ausschalten, danach wieder einschalten Siehe auch Kapitel „Bus-Statusanzeige“ auf Seite 34. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Byte 8 gerätespezifische Erweiterungen weitere gerätespezifische Diagnose Byte 241 (max) TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 60 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Page 61
Überlauf bei erweiterter Bit 7 Ext_Diag_Overflow Diagnose Bit 6-0 Reserviert TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 61 of 139...
Stehen zusätzliche Diagnoseinformationen zur Verfügung, so trägt der Slave an dieser Stelle die Anzahl der Bytes ein, die außer der Normdiagnose noch folgen. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 62 of 139...
Bit 6 nicht benutzt Bit 7 nicht benutzt TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 63 of 139...
Für zusätzliche Alarme ist das Byte 17 reserviert, jedoch sind keine weiteren Alarme implementiert. Erweiterte Diagnose, Byte 17 Bedeutung Bit 0-7 reserviert TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 64 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Erweiterte Diagnose, Byte 24-25 Byte 24 Revisions-Nummer Byte 25 Revisions-Index TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 65 of 139...
Erweiterte Diagnose, Byte 48-57 Die Diagnosebytes zeigen Seriennummer des Encoders an. Wird diese Funktion nicht unterstützt, werden Sterne angezeigt (Hex-Code 0x2A) **********. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 66 of 139...
Derzeit wird im Profil nur der Alarm "Positionsdaten-Fehler" und die Warnung "zul. Temperatur überschritten" unterstützt. Sonstige Warnungen sind nicht frei geschaltet und werden auf die im Profil vorgesehenen Standardwerte gesetzt. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016...
Wird der 15-pol. SUB-D Stecker nicht benötigt, schützt diese Abdeckung den 15-pol. SUB-D Stecker vor Verschmutzungen. Bestellnummer Beschreibung 49.500.041 Steckerabdeckung IP65 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 68 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
99-1437-810-05 PROFIBUS_OUT, Kabelstecker 5-pol. M12x1, B-kodiert 99-0436-14-05 Versorgung, Kabeldose 5-pol. M12x1, A-kodiert TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 69 of 139...
Page 70
Zubehör TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 70 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Page 71
User Manual Laser Measuring Device LLB-65 / LLB-500 PROFIBUS-DP TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 71 of 139...
Page 72
PROFIBUS-DP and the PROFIBUS logo are registered trademarks of PROFIBUS Nutzerorganisation e.V. (PNO) [PROFIBUS User Organization] SIMATIC is a registered trademark of SIEMENS corporation TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 72 of 139...
5.1 PROFIBUS-DP interface ......................92 5.1.1 DP Communication protocol ................... 92 5.2 Analog interface ........................93 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Page 74
8.5.12 Error indicating current ..................123 8.6 Configuration example, SIMATIC Manager V5.3 ..............124 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 74 of 139 TR - ELE - BA - DGB - 0022 - 04...
Page 75
10.3 Laser glasses ........................138 10.4 Connector cover IP65 ......................138 10.5 PROFIBUS / Supply – Mating connector ................139 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
– New design 02/25/15 – Laser lifetime Reference to Support-DVD removed 02/08/16 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 76 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
the operator's operating instructions specific to the system, this User Manual TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 77 of 139...
1.2 EC Declaration of conformity The measuring systems have been developed, designed and manufactured under observation of the applicable international and European standards and directives. A corresponding declaration of conformity can be requested from TR-Electronic GmbH. The manufacturer of the product, TR-Electronic GmbH in D-78647 Trossingen, operates a certified quality assurance system in accordance with ISO 9001.
EC EMC Directive, the harmonized standards, European standards or the corresponding national standards. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016...
observing the operating instructions from the machine or system manufacturer, operating the measuring system within the limit values specified in the technical data (Assembly Instruction / User Manual). TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 80 of 139...
Repairs carried out autonomously. Third party interference and Acts of God. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 81 of 139...
Define clear rules of responsibilities for the assembly, installation, start-up and operation. The obligation exists to provide supervision for trainee personnel ! TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 82 of 139...
Observe the legal and local regulations applicable to the operation of laser units. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Page 84
Clean the lens opening of the measuring system regularly with a damp cloth. Do not use any aggressive detergents, such as thinners or acetone! TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 84 of 139...
Figure 1: Standard application Key features: ● LLB-65/LLB-500 measurement range on natural surfaces 0.05 m up to approx. 65 m ● LLB-500 measurement range on (reflective) target plate 0.5 m up to approx. 500 m ● PROFIBUS-DP interface ●...
Receiver optics Product label, see 2.9 on page 83. 3x M12 male/female connector, PROFIBUS-DP – interface, Supply TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 86 of 139 TR - ELE - BA - DGB - 0022 - 04...
(e.g. > 150m at LLB-500). These effects are described in: B.Edlen: “The Refractive Index of Air, Metrologia 2”, 71-80 (1966) TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016...
Erroneous measurements can occur in the case if the laser beam is reflected from other objects than the target. Avoid any reflecting object along the measurement path. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany...
953 mbar air pressure. The influences of the atmospheric environment are described in B.Edlen: “The Refractive Index of Air, Metrologia 2”, 71-80 (1966) TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
1 digital output for error status See 4.1 Measuring accuracy on page 89. In case of permanent continuous measurement the max. temperature is reduced to 45°C TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 90 of 139...
4.3 Physical dimensions All dimensions in mm TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 91 of 139...
The basic functionality is defined by the performance level V0. This includes cyclic data exchange, as well as the station, module and channel-specific diagnosis. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 92 of 139...
The analog output of the LLB is a current source (0…20 mA or 4…20 mA). It is capable of driving loads up to 500 The analog output at LLB-65 has an accuracy of ± 0.2% full scale and at LLB-500 an accuracy of ± 0.1% full scale.
For trouble-free operation use a separate power supply for the For the supply voltage a cable cross section of min. 0.75 mm is recommended. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 94 of 139...
Configuration see page 120 and 121. DOUT Figure 3: Wiring of the Digital Outputs TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
In particular, the applicable EMC directive and the shielding and grounding guidelines must be observed! TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 96 of 139...
= this state, subsequent PROFIBUS is decoupled. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 97 of 139...
Make sure, that the total resistance in the analog path is lower than 500 Configuration see page 119, 122 and 123. Figure 4: Connection of an analog instrument and a PLC TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 98 of 139...
Analog Ground Analog Output 0…20 mA or 4…20 mA Device Ground Device Ground TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 99 of 139...
Analog ground (AGND). GND and AGND are connected to the housing by a RC element, see Figure 5. Figure 5: Connection between shield, Ground (GND) and Analog Ground (AGND) TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 100 of 139...
POWER (green): ON, if the voltage supply is in the permissible range Figure 6: Device Status LEDs TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
This achieves a high level of security against projection errors. The measuring system has the ID number 0D65 (hex). TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 102 of 139...
Return Inputs WCFG = Wait Configuration DXCHG = Data Exchange Figure 8: DP slave initialization TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Non-recoverable measuring system fault Corresponding measures in case of an error see chapter “Diagnosis over optical bus- status-display”, page 129. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 104 of 139 TR - ELE - BA - DGB - 0022 - 04...
0 = Consistency about one byte or word Consistency: 1 = Consistency about the complete module TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
- Min current - Error indicating current from the bus master perspective TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 106 of 139 TR - ELE - BA - DGB - 0022 - 04...
3 – 0 Data Consistency Word format Input data Length code TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 107 of 139...
Page 108
– 2 Data Default (dec.) 0 (–19685040 - +19685040) Offset value TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 108 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Page 109
Data Default (dec.) 0 (–19685040 - +19685040) Error output value TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 109 of 139...
3 – 0 Data Consistency Word format Input data Length code TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 110 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Page 111
1/10 mm (Default) MIL (1/1000 Inch) 1/100 Inch 1/10 Inch Inch TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 111 of 139...
Page 112
Error value Min current 0 mA 4 mA TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 112 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Page 113
Data Default (dec.) 0 (–19685040 - +19685040) Digital Output 1 On TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 113 of 139...
Page 114
Data Default (dec.) 0 (–19685040 - +19685040) Digital Output 2 Off TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 114 of 139 TR - ELE - BA - DGB - 0022 - 04...
Page 115
Data 0 (0 – 200) Default (dec.) Error indicating current TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 115 of 139...
Lower limit Upper limit according to the adjusted resolution, related to 500 m TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 116 of 139 TR - ELE - BA - DGB - 0022 - 04...
Upper limit 0x01 2C 5E B0 (+19 685 040) Default 0x00 00 00 00 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
0 adjusts automatically the fastest sampling rate. The measurement can need at LLB-65 0.15 up to approx. 4 s and at LLB-500 0.04 up to approx. 4 s depends on the surface condition. The value 0xFF deactivates this function and the laser diode is switched off.
Current range: 0…20 mA 0 mA Current range: 4…20 mA 4 mA TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 119 of 139...
For that purpose the parameter “Measuring cycle” must be adjusted to the value “255” decimal. The distance measuring is performed automatically as fast as possible. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 120 of 139...
Upper limit 0x01 2C 5E B0 (+19 685 040) 0x00 00 00 00 Default TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Offset or Preset. The operating range begins at the measuring reference, see also page 91 and ends at LLB-65 with 65 m or at LLB-500 with 500 m. In case of breach of condition, an incorrect current value will be output! ...
Page 110 Lower limit 0x00 Upper limit 0xC8 Default 0x00 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 123 of 139...
File names and entries in the following masks are to be regarded only as examples of the procedure. For the GSD file to be transferred to the catalogue, it must first be installed: TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 124 of 139...
Page 125
110 The entry Universal module is erroneously available for some systems, but must not be used! TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Page 126
Once the measuring system is connected to the master system, the network settings can be undertaken --> Object Properties... --> PROFIBUS... button): TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 126 of 139...
Page 127
Transfer the required configuration from the catalogue to the slot (drag&drop). The measuring system symbol must be active. Perform parameterization with a double click on the slot number: TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016...
– Replace measuring system – Hardware failure See also chapter “Device Status display” on page 101. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 128 of 139 TR - ELE - BA - DGB - 0022 - 04...
- Analogue output min and max – Switch-off device, switch-on again See also chapter “Bus status display” on page 104. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
8 device-specific extensions further device-specific diagnosis byte 241 (max) TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 130 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
7 Ext_Diag_Overflow Overrun for extended diagnosis bit 6-0 Reserved TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 131 of 139...
If further diagnosis informations are available, the slave enters the number of bytes at this location, which follow in addition to the standard diagnosis. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 132 of 139...
6 not used bit 7 not used TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 133 of 139...
Byte 17 is reserved for additional alarms, however no further alarms are implemented. Extended diagnosis, byte 17 Significance bit 0-7 reserved TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 134 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
Extended diagnosis, bytes 24-25 byte 24 Revision number byte 25 Revision index TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04 Page 135 of 139...
The diagnosis bytes show the serial number of the encoder. If this function is not supported, asterisks ********** (hex code 0x2A) are displayed. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 136 of 139...
Further warnings aren't available and will be set to the default values prescribed by the profile. TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
If the 15 pin D-Sub connector is not used, this cover protects the 15 pin D-Sub connector. Part Number Description 49.500.041 Connector cover IP65 TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany Page 138 of 139 TR - ELE - BA - DGB - 0022 - 04 02/08/2016...
PROFIBUS_OUT, male connector 5-pol. M12x1, B coded 99-0436-14-05 Supply Voltage, female connector 5-pol. M12x1, A coded TR-Electronic GmbH 2011, All Rights Reserved Printed in the Federal Republic of Germany 02/08/2016 TR - ELE - BA - DGB - 0022 - 04...
Need help?
Do you have a question about the LLB-65 and is the answer not in the manual?
Questions and answers